CONCORD 75023 Operator's Manual page 31

Handheld pulse oximeter
Table of Contents

Advertisement

Page30
75023
Note 1: From 80 MHz to 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a: Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the device.
b: Over the frequency range from 1 50kHz to 80MHz, field strengths should be
less than 3V/m.
Recommended Se paration Dletancae between Portab l e end Mob l l e RF
The device is suitable for use in an electromagnetic environment in which
Com m uni cation■ E o u l o ment end The device
radiated RF disturbance are controlled. The customer or the user of the
device can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications
equipment (transmitters) and the device as recommended below, according
to the maximum output power of the communication equipment.
Rated Maxi m u m
Output power of
Trane m l tte r (W)
0 . 0 1
0 .1
1
1 0
For transmitters at a maximum output power not listed above, the separation
1 00
distance can be estimated using the equation in the corresponding column,
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1: From 80 MHz to 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Se paration Dletance (m) Corree pondlng to F re q uency
of Trane m l tte r
1 50k to 80MHz
#
d = 1 . 1 6
0.1 2
0.38
1 .2
3. 8
1 2
Copyright. Concord Health Supply. Inc.
SO M to 800MHz
#
d = 1 . 1 6
0. 1 2
0.38
1 .2
3.8
1 2
BOOM to 2. 5GHz
d = 2 33
✓P
0.23
0.73
2.3
7.3
23