pulse rate low alarm limit is limited one number lower than the pulse rate
upper alarm limit. The pulse rate low alarm limit cannot be set equal to or
higher than the pulse rate upper alarm limit. In the same way, the pulse
rate upper alarm limit cannot be equal to or lower than the pulse rate low
alarm limit.
Press button [SET]---->[SPO2 L] and Press the Up/Down button, set the

5.3.3.2 Set SPO2 Low Alarm Limit

SPO2 value between 20 and 100.
Note:The %SpO2 alarm limit range is 20% to 100%. The upper value of
the %SpO2 low alarm limit is limited to the %SpO2 upper alarm limit. The
%SpO2 low alarm limit cannot be set equal to or higher than the %SpO2
upper alarm limit. In the same way,the lower value of the %SpO2 upper
alarmlimit is limited to the %SpO2 low alarm limit. The %SpO2 upper
alarm limit cannot be set equal to or lower than the %SpO2 low alarm limit.
Select [SET]-->[Aln], and press the Up/Down button, to change the

5.3.3.3 Switch ON/OFF Alarm Volume

ala,m ,oi,me. Th
WARNING, When the alarm sound is switched off, the pulse oximeter
will give no audible notice tones even if a new notice occurs. Therefore
A
the user should be very careful about whether to switch off the notice
sound or not.
WARNING, Do not rely exclusively on the audible notice system for
user monitoring. Adjusting alarm volume to a low level may result in a
A
hazard to the user. Always keep the user under close surveillance.
Press the /$ key for 1 second, the /$ key will light and the alarm
5.3.3.4 Alarm Silencing
messages remain displayed. The alarm will stay silent until the)$ key is
pressed again.
A_ WARNING, The 75023 will only keep the last setting. When the 75023
� is turned on, user should check whether the current selling meets
conditions of user.
When an alarm occurs, observe the following steps to take proper actions:

5.3.4 When an Alarm Occurs

1) Check the user's condition.
2) Confirm the alarm parameter or alarm category.
3) Identify the source of the alarm.
4) Take proper action to eliminate the alarm condition.
Copyright. Concord Health Supply. Inc.
� = r � f � r
f
". The h;ghest '""' ;, 10.
f
Page 15