75023
To access the following parameters below, press the set button until

5.3 Set up the 75023

the parameter you want to adjust is displayed.
When you have completed adjusting the parameters, press and
release the power button once, to save your parameters and return to
the main screen. Please note that if you do not press the power button at
the conclusion of this process, your parameters will not be saved.

5.3.1 Adjust Pulse Beep Volume

1) Press button [SET]-->[VOL] and press Up/Down button to increase or
decrease the pulse beep volume, then set the volume between O and 10.
Note, When the pulse beep volume display is shown, the user is able to
adjust the volume of the pulse beep tone. Each activation of an Up or
Down button increases or decreases the pulse beep volume and
increments or decrements by one the number of segments shown on the
Pulse Amplitude (blip) bar as a relative indicator of the current volume.
The minimum pulse rate volume is none or OFF (no blip bar segments
shown), the maximum pulse rate volume is ten (ten segments).
Attempted adjustments outside the range generate an invalid key tone.
2) Press power button for 1 second, the menu screen can quickly return to
normal operation display.

5.3.2 Select User Type

WARNING, Be sure to select correct user category setting for your
� user before measurement. Wrong user category may result in user
A
hazard due to pulse rate alarm limits.
1) Press button [SET]-->[PA] and Press the Up/Down button, select
user type value: A DU(adult), PED(pediatric), NEO(neonate).
2) Press the power button for 1 second, the menu screen can quickly
return to normal operation display.

5.3.3 Set Alarm Limit

Press button [SET]-->[BPM HJ or [BPM L]and press the Up/Down button,

5.3.3.1 PR High/Low Alarm Limit

set the PR value between 30 and 250.
Note: The pulse rate alarm limit range is 25 to 250.The upper value of the