User Manual WS Series Pulse Oximeter Please read the manual carefully before use...
Page 4
Introduction Thank you for your purchase of our pulse oximeter (“oximeter”). Prior to use of the product, please read the content of this manual carefully to ensure proper use of the product. After reading, please keep this manual properly for future reference. Hunan Accurate Bio-Medical Technology Co., Ltd.
Page 5
Foreword Introduction This manual introduces in detail the use and functions of the product ● as well as how to operate it. Prior to use of the product, please carefully read and understand the content of this manual to ensure proper use of the product and safety of user.
1.Safety 1.1 Safety Information DANGER Indicates an imminently hazardous situation, which, if not avoided, ● could result in death, serious injury or property damage. WARNING Alerts you to potential dangers or unsafe operations, which, if not ● avoided, may result in death or serious injury or property damage. CAUTION Alerts you to potential dangers or unsafe operations, which, if not ●...
accessories to prevent the patient from getting wound or suffocated, entanglement of the cables, or electrical interference. Only use the SpO2 probe supplied by the manufacturer; use of a ● SpO2 probe from other source could result in performance degradation or damage of the device or cause safety risks.
following factors could degrade the measurement performance and accuracy of the oximeter: 1)Strong interference by light (e.g., fluorescent light, dual ruby light, infrared heater, operating light, direct sunlight) in the application environment will affect the measurement accuracy. 2)Water vapor and mist in the device. 3)The size of finger measured is out of range.
Do not attempt to open the housing for repair. If the product is ● damaged and needs repair, it can be repaired by qualified service Personnel designated by the manufacturer. The manufacturer may provide the service personnel with the Service Manual which contains information necessary for repair such as circuit diagram, component list, legend and correction rules.
2.Overview 2.1 Introduction 2.1.1 Scope of Application The product is intended for monitoring of user’s SpO2 and PR in hospitals and at home. WARNING ● The oximeter should be used by or under the guidance of medical workers. When using the device at home, user should carefully read the User Manual before use and where necessary, consult the doctor, dealer or manufacturer.Human contact part of the equipment meet the bio-compatibility requirements and complies with ISO 10993-1, ISO...
1.Display screen Display the SpO2 and SpO2 measurement trend graph; PR value, PR measurement trend graph, pulse intensity bar chart, perfusion index (PI) value, pulse wave, operating state of main unit, measurement duration, user ID, battery level, and time. 2. Power/Confirm key ■...
3.Preparation before Use 3.1 Unpacking Inspection Please check the packing box carefully before opening it. Please get in touch with the carrier immediately if any damage is found. Properly open the packing box; carefully take the device and other components out of the packaging box, and count them item by item according to the Packing List.
3.3 Start up 1.Prior to start up, please check whether the oximeter has any mechanical damage. 2. Make sure the remaining battery capacity is adequate. 3. Press the Power key to enter the main interface. 3.4 Shut down Please shut down the oximeter according to the following steps: 1.
Page 16
Set Intended Population ● 1.Press the Confirm key to confirm the selection of Intended Population; 2. Press the Up/Down key for selection; 3. Press the Confirm key to confirm the modification and return to other menu for selection. Reminder ON/OFF ●...
4.2 System Setup Time Setup (disabled when the main unit is inserted with a probe) ● 1. Measurement interface → Press the Confirm key to enter the interface → Squared menu → System Setup → Time Setup; 2. Press the Up/Down key to select the setting item, and press the Confirm key to confirm the selection;...
and know the measured results. Note: The report is for reference only and cannot be used as a basis for treatment. 4.5 Battery Level Detection The device automatically monitors the battery level, and displays and updates it on the display screen. When the battery level is displayed as , please charge the battery timely.
5.SpO2 Measurement and Information 5.1 Overview Continuous non-invasive pulse SpO2 oximetry is employed for SpO2 measurement. It measures the luminous flux of light of specific wavelength emitted by the luminous light source of SpO2 probe after absorption by oxyhemoglobin in the patient’s tissue and arrival at the photoelectric detector, thus to obtain SpO2 and PR.
6. Bar chart: The amplitude of bar chart represents the level of pulse strength. 7. PI: Perfusion index. 8. Measurement duration: Records the duration of measurement. 9. Main unit operating state: The current state of the main unit. 10. PR measurement trend graph 11.
since blood flow occlusion during BP measurement will affect the SpO2 reading. 5.3 Measurement Steps 1. Clean the measuring position, such as colored nail polish. 2. Place the SpO2 probe at the measuring position. 3. Connect the main unit and SpO2 probe. 4.
6.1.1 Lithium Battery Charging Operation steps: 1. Connect the charging cable to the multi-purpose multiplex interface of the oximeter; 2. Connect the other end of the charging cable to the charger; 3. Disconnect the adapter after full charging. ▲ NOTE The service life of lithium battery depends on the time and frequency ●...
Never soak the device in any liquid. ● Never pour any liquid onto the device or its accessories. ● Never allow any liquid to flow into the housing. ● Never use abrasive materials (e.g., steel wool or silver polish) or ●...
2. Air-dry the device or use a dry, clean cloth to wipe the surface. Recommended period: After each use of the device. WARNING: Do not use high-temperature and high-pressure gas to disinfect the device. 7.3 Scrapping To avoid contaminating the environment or other equipment or infecting other people, please disinfect and purify the device and its accessories according to applicable national laws or regulations before scrapping, and also observe the local regulations on scrapping of medical wastes.
A.Product Specifications Type Wavelength Power 660±6nm 1.8mW 905±10nm 2.0mW The range of emission wavelength is 600~1000nm; information on the wavelength range could be especially useful for clinical physicians. Safety specification (classification according to IEC 60601-1) Type of protection Class II device powered by an internal electrical against electrical shock power source Rating of protection...
Page 26
Lithium battery Quantity Specification 3.7V Battery capacity 500mAh Voltage supply time When the device performs measurement continuously after the lithium battery is fully charged, the working time should not be less than8h (under the conditions of long-time measurement, screen off and no measurement abnormality) Environmental Operation...
Technical Description The table below shows the statistical conclusion of the study on invasive controlled desaturation according to Annex EE “Guideline for evaluating and documenting SpO2 accuracy in human objects” of ISO 80601-2-61. The statistical result shows the accuracy distribution within the range of 70%~100%, which is helpful for user.
Page 28
B.1 PC management software for the pulse oximeter: The operating ● environment is a PC installed with WIN7/WIN10 OS or higher compatible version as well as Anti-virus software. Minimum hardware configuration requirements: CPU: 1GHz or faster; memory: 1GB or larger; hard disk: >20GB; graphics: >128MB;...
management software and the mobile phone management software for the pulse oximeter; user types include Super Admin, Admin and Ordinary User. ● Super Admin: Create and delete the Admin account and Ordinary User account; view operating parameters and measured results. ●...
Appendix Guidance and Manufacturer’s Statement - Electromagnetic Emissions The pulse oximeter is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it is used in such an environment: Emission test Compliance Electromagnetic environment – guidance The pulse oximeter uses RF energy only for its internal function.
Page 31
IEC 60601 Compliance Electromagnetic Immunity test test level level environment – guidance ± 1 kV line to line Surge ± 2 kV line to IEC 61000-4-5 ground < 5% UT (> 95% dip Voltage dips, short in UT) for 0.5 cycle interruptions and 40% UT (60% dip in voltage variations...
Page 32
In the above formula: P – the transmitter’s maximum rated output power (W) learnt from the transmitter manufacturer; d – the recommended separation distance (m)b. The field intensity of fixed RF transmitters is measured by surveyingc the electromag- netic field and should be lower than the compliance level in either frequency ranged. This product may cause interference to the nearby equipment marked with the following symbol: Note 1: At 80 MHz and 800 MHz, the higher frequency band applies.
For transmitters’ maximum rated output power not listed above, the recommended separation distance “d” in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where “P” is the transmitter maximum rated output power in watts (W) provided by the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the higher frequency band applies.