Physiological Notification Messages; Technical Notification Messages - CONCORD 75023 Operator's Manual

Handheld pulse oximeter
Table of Contents

Advertisement

75023
Noti ce
Notifi cation
SpO2 Too High
SPO2 reading will flash.
M■■■■ae
Notification lamp is red
and flashing.
SpO2 Too Low
SPO2 reading will flash.
Notification lamp is red
and flashing.
PR Too High
PR reading will flash.
Notification lamp is red
and flashing.
PR reading will flash.
PR Too Low
Notification lamp is red
and flashing.
No Pulse
!;' R � nd SPO2 reading is
Notification lamp is red
- -
and flashing.
Low Perfusion
Notification Lamp
remains yellow. The
screen will show " Low
Perfusion "
Low signal
Notification lamp is
yellow and flashing. The
Quality
screen will show " Low
Signal Quality"
Noti ce
Notification
Me■e■oe
Notification lamp is
yellow and flashing.
Sensor Off
The screen will show
"OFF"
Page 20
Physiolog ical Notification Messages
Technical Notification Messag es
Notl c1 Cau■■ and action■
H
A measurement has risen
Level
above the high alarm limit
or fallen below the low notice
limit. Check the person's
H
condition and check if the
fi erson category and notice
mil settings are correct.
H
H
H
The pulse signal was too
weak to be analyzed. Check
the person's condition,
SpO2 sensor and
measurement site.
The pulse signal was weak
L
and please Check the user's
condition, check SpO2
sensor and measurement
site.
The pulse signal was too
weak , Check the person's
condition, SpO2 sensor and
M
measurement site.
NOIIC!I C■u■e and action■
The SpO2 sensor detached
Level
from the person the pulse
oximeter, or there was a fault
with the SpO2 sensor, or an
unspecified SpO2 sensor was
used. Check that the sensor
application site and the sensor
type are correct, and make
sure that the sensor is
M
undamaged.
Reconnect the sensor if the
sensor is disconnected or use a
new sensor if the sensor is
damaged, for example.