Question
How can I see all the current settings for
my monitor?
I have made the monitor configuration
changes I want, but they are not being
stored. Why?
How can I change the analog output
voltages from Corometrics standard to
Hewlett-Packard / Phillips standard?
Revision D
Troubleshooting: FAQs
1. Enter the Install Options Screen 1.
options screens?"
on page 5-27.)
2. In the Install Options Screen 1, rotate the Trim Knob to highlight the PrintAll softkey
at the bottom of the display, below the menu bar. This initiates the PrintAll function.
3. The display will change to one of the setup screens, and the printer will start. The
printer will output a hardcopy version of the setup screen that is displayed.
4. After the first setup screen prints, the display switches to the next setup screen, and
again the printer outputs a hardcopy version of this screen.
5. PrintAll will output in order, recopies of the following setup screens:
Setup (256 and 259 models)
FECG or US (256 and 259 models)
NIBP Setup (259 model only)
MHR/P Setup (259 model only)
MSpO
Setup (259 model only)
2
Vital Signs History (256 and 259 models)
Master Alarm Setup (259 model only)
General Setup (256 and 259 models)
Install Options Screen 1 (256 and 259 models)
Install Options Screen 2 (256 and 259 models)
Communications Setup (256 and 259 models)
Diagnostic Control (256 and 259 models)
6. Wait until the printer stops and the display returns to the Install Options Screen 1
before you exit the Service mode.
7. To exit the Service mode following the instructions at the top of
Refer to
"How are the 250 configurations settings stored and recalled?"
1. The analog voltage outputs at J102 may conform to one of two voltage standards for
analog central station outputs: Corometrics or Hewlett-Packard. The default is
Corometrics.
2. To change the setting, remove the unit cover.
3. At the rear of the unit, on the main CPU motherboard, below the communications
board, there is an 8 bit DIP switch, SW1.
4. Switch number 2 determines the analog output standard used by the 250 Series
Monitor.
If switch number 2 is on (closed), the output voltages comply with the
Corometrics standard.
If switch number 2 is off (open), the output voltages comply with the Hewlett-
Packard Standard.
250 Series Maternal/Fetal Monitor
2020551-001
Answer
("How do I access the service setup / install
Page 16.
on page 5-16.
5-35