Affirm Breast Biopsy Guidance System Service Manual
Appendix D: Biopsy Device Evaluation Procedure
Figure 51: Biopsy Device Parameters - Front Edge for Lateral Arm
Figure 52: Biopsy Device Parameters - Body Size for Lateral Arm
D.2.2

Needle Validation Process

Note
All manual jogs in the following procedure are performed from the Biopsy Control
Module Target Screen, not from the Service Screen. (This screen affects the target
numbers.) The Jog is used to move the needle (and target) away from the phantom
tip, which prevents a collision of the needle tip and phantom tip. This method also
conveniently provides intuitive results without having to correct for the Jog
movement.
Verify the System is Correct for a Known Needle
1.
2.
3.
4.
Page 82
Sign on to the Acquisition Workstation as a Service User.
Install a biopsy Compression Paddle.
Position the Targeting Phantom on the breast tray and center it in the Biopsy Paddle
window (does not have to be exact).
Compress the Targeting Phantom with ~10 pounds/~45 newtons of force.
IMPORTANT: Do not move the phantom for the remainder of this procedure.
MAN-05754 Revision 002