6.1.4. The optimal position is at the average depth of measured material, it will generally locate at
mountainside between the peak and bottom (the cone angle from by the pouring process), indicates
below.

6.2. Installation Instruction

6.2.1. During installation, the flange should be mounted at horizontal. Besides, the housing and cable wire
should be kept in vertical direction related to the measured material level. It should be carefully
checked that the flange can let the wire cable move free and doesn't rub against the body.
6.2.2. On demand, user may connect an extending tube to connect the flange. If you do that, keep it in mind
that minimum diameter should not be less than 4".
6.2.3. For leakage, FineTek suggests customer should use O-ring seal or washer between the flange
connections and secure it if indeed.
6.3. Caution
6.3.1. During installation, user should carefully check that the cable wire is wound up well in pulley set and
no kinks, broken, or any abnormal compress on the cable wire.
6.3.2. The cable wire should be put on the hole of the weight head connector and be secured tightly by
screwdriver.
6.3.3. Firmly secure the screws to fix the front cover and body, otherwise the dust or powder will pour into
9