7. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
8. When the correct pressure is reached, the pumping stops and the
pressure falls gradually. If the required pressure was not reached,
the device will automatically pump some more air into the cuff.
9. During the measurement, the pulse indicator BT flashes in the
display.
10.The result, comprising the systolic AM and the diastolic AN
blood pressure and the pulse rate AO is displayed. Note also the
explanations on further display symbols in this booklet.
11.When the device has finished measuring, remove the cuff.
12.Switch off the device. (The monitor does switch off automati-
cally after approx. 1 min.).
How not to store a reading
As soon as the reading is displayed press and hold the ON/OFF
button 1 until «M» AS is flashing. Confirm to delete the reading by
pressing the M-button 2.
You can stop the measurement at any time by pressing the
ON/OFF button (e.g. if you feel uneasy or an unpleasant
pressure sensation).
If the systolic blood pressure is known to be very high,
it can be an advantage to set the pressure individually. Press
the ON/OFF button after the monitor has been pumped up to
a level of approx. 30 mmHg (shown on the display). Keep the
button pressed until the pressure is about 40 mmHg above
the expected systolic value – then release the button.
5. Appearance of the Atrial Fibrillation Indicator for
early Detection (Active only in AFIB/MAM mode)
This device is able to detect atrial fibrillation (AF). This symbol AP
indicates that atrial fibrillation was detected during the measure-
ment. Please refer to the next paragraph for information regarding
the consultation with your doctor.
BP A200 AFIB
Information for the doctor on frequent appearance of the
atrial fibrillation indicator
This device is an oscillometric blood pressure monitor that also
analyses pulse irregularity during measurement. The device is
clinically tested.
The AFIB symbol is displayed after the measurement, if atrial fibrilla-
tion occurred during measuring. If the AFIB symbol appears after
having performed a full blood pressure measurement episode (tripli-
cate measurements), the patient is advised to perform another
measurement episode (triplicate measurements). If the AFIB symbol
appears again, we recommend the patient to seek medical advice.
If the AFIB-symbol appears on the screen of the blood pressure
monitor, it indicates the possible presence of atrial fibrillation.
The atrial fibrillation diagnosis however, must be made by a
cardiologist based on ECG interpretation.
In the presence of atrial fibrillation the diastolic blood pressure
value may not be accurate.
In the presence of atrial fibrillation using MAM-mode is recom-
mended for more reliable blood pressure measurement.
Keep the arm still during measuring to avoid false readings.
This device may not or wrongly detect atrial fibrillation in
people with pacemakers or defibrillators.
6. PC-Link Functions
This device can be used in conjunction with a personal computer
(PC) running the Microlife Blood Pressure Analyzer+ (BPA+) soft-
ware. The memory data can be transferred to the PC by
connecting the monitor via a cable.
If no CD and cable is included download the BPA+ software from
www.microlife.com/software and use a USB cable with a Mini-B 5
pin connector.
Installation and data transmission
1. Insert CD into the CD ROM drive of your PC. The installation
will start automatically. If not, please click on «SETUP.EXE».
2. Connect the monitor via the cable to the PC; there is no need to
switch the device on. 3 horizontal bars will appear on the
display and last for 3 seconds.
3. The bars will then flash to indicate that the connection between
PC and device is successful. As long as the cable is plugged in,
the bars will keep flashing and the buttons are disabled.
EN
5

Advertisement

Table of Contents
loading

Table of Contents