Pfeiffer Vacuum XPT 200 PB Supplementary Information

Pfeiffer Vacuum XPT 200 PB Supplementary Information

Digiline gauge with profibus interface

Advertisement

Quick Links

SUPPLEMENTARY INFORMATION
XPT 200 PB ∣ CCT 36X PB ∣ CCT 37X PB
DigiLine gauge with Profibus interface
RS485
Profibus
device status
status
1
0
0
RS485
15
3
2
2
adr
12
4
4
13
5
10
6
6
11
7
8
8
9
x10
address
x1
EN
Translation of the Original

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the XPT 200 PB and is the answer not in the manual?

Questions and answers

Summary of Contents for Pfeiffer Vacuum XPT 200 PB

  • Page 1 RS485 Profibus device status status RS485 address SUPPLEMENTARY INFORMATION Translation of the Original XPT 200 PB ∣ CCT 36X PB ∣ CCT 37X PB DigiLine gauge with Profibus interface...
  • Page 2 Dear Customer, Thank you for choosing a Pfeiffer Vacuum product. Your new gauge is designed to support you in your individual application with maximum performance and without malfunctions. The name Pfeiffer Vacuum stands for high-quality vacuum technology, a comprehensive and complete range of top-quality products and first-class service.
  • Page 3: Table Of Contents

    Table of contents Table of contents About this manual Validity 1.1.1 Applicable documents 1.1.2 Variants Conventions 1.2.1 Abbreviations Trademark proof Product description Function 2.1.1 Structure of the CCT gauge 2.1.2 Structure of the xPT gauge Installation Establish Profibus connection Configure Profibus connection for CCT gauges Configure Profibus connection for xPT gauges Profibus modules Input data...
  • Page 4: About This Manual

    ● DigiLine gauge with Profibus interface The part number is found on the rating plate of the product. Pfeiffer Vacuum reserves the right to make technical changes without prior notification. Information that relates to only one of the products is indicated accordingly.
  • Page 5: Trademark Proof

    About this manual Abbreviation Explanation Ground Device master data (Profibus) High vacuum IEEE Institute of Electrical and Electronics Engineers Cold cathode Overrange Pirani Profibus DP Profibus (process field bus) – non-local peripherals Totally integrated automation Underrange Tbl. 2: Abbreviations used 1.3 Trademark proof ●...
  • Page 6: Product Description

    Product description 2 Product description 2.1 Function It is possible to connect to a Profibus-DP system via the connector (M12, B-coded) designated "Profi- bus". The interface is galvanically safe and is isolated from the maximum supply voltage. The gauge's voltage supply is provided via the RS-485 connector. 2.1.1 Structure of the CCT gauge RS485 device...
  • Page 7: Installation

    3. Establish the Profibus connection in compliance with the applicable guidelines. 4. Connect the voltage supply to the gauge via the RS-485 connector in compliance with the regula- tions in force. 5. Use a suitable connection cable and components from the Pfeiffer Vacuum line of accessories. 7/24...
  • Page 8: Configure Profibus Connection For Cct Gauges

    Installation 3.2 Configure Profibus connection for CCT gauges Profibus configuration tools There are various suppliers of configuration tools for configuring the Profibus. The configu- ration procedure is identical. The figures show the Siemens TIA portal by way of an exam- ple. Fig.
  • Page 9 Installation Fig. 5: Drag & drop input and output modules into slots1 to 7. Fig. 6: Integrate input and output modules 9/24...
  • Page 10 2. Open the configuration tool. 3. Import the unzipped files (GSD file and image) into the process control. – You will find the CCT gauge under the path PROFIBUS DP/Gateways/Pfeiffer Vacuum/CCT 3xx PB. 4. Drag and drop CCT 3xx PB into the configuration tool's network area and integrate CCT 3xx PB there.
  • Page 11: Configure Profibus Connection For Xpt Gauges

    Installation 3.3 Configure Profibus connection for xPT gauges Profibus configuration tools There are various suppliers of configuration tools for configuring the Profibus. The configu- ration procedure is identical. The figures show the Siemens TIA portal by way of an exam- ple. Configure switching ranges for xPT gauges You can configure the switching ranges in the Profibus parameter data for the HPT 200 PB, MPT 200 PB and RPT 200 PB gauges.
  • Page 12 Installation Fig. 9: Drag & drop gauge type for inputs and outputs into slot 1 of modules Fig. 10: Integrate input and output modules on TIA portal 12/24...
  • Page 13 – You will find the xPT gauge under the path PROFIBUS DP/Gateways/Pfeiffer Vacuum/xPT 200 PB. 4. Drag and drop xPT 200 PB into the configuration tool's network area and integrate xPT 200 PB there. 5. Select the input and output modules for the gauges you are using.
  • Page 14 Installation 7. Compile the entries. 8. Check the correct function of the entries online. 14/24...
  • Page 15: Profibus Modules

    Profibus modules 4 Profibus modules 4.1 Input data Byte Meaning Gauges Actual pressure Actual pressure value in IEEE 754 format (floating value point) Status bits Degas active or filament switched off Value range infringement pressure value in output data Command supported in output data Value range infringement Pi correction factor in output data Value range infringement BA/CC correction factor...
  • Page 16: Diagnostics Data

    Profibus modules Compute effective correction factor The admissible value range of the entered correction factors is between 1 and 800. It is equivalent to effective correction factors between 0.01 and 8.00. ► Multiply the entered correction factor by 0.01 to calculate the effective correction factor. 4.3 Diagnostics data Byte Meaning...
  • Page 17: Operation

    Operation 5 Operation 5.1 Operate gauge via Profibus interface NOTICE Data transmission error due to simultaneous operation on both interfaces If you attempt to operate the gauge simultaneously via the RS-485 and Profibus interfaces, this will result in incorrect data and interference with the data transmission. ►...
  • Page 18: Profibus Operating Mode Display Via Led

    Operation 5.2 Profibus operating mode display via LED LED status Display Meaning xPT gauge CCT gauge Profibus on device side not ac- Profibus on device side not ac- tive tive No Baud rate detected Baud rate detected, no user data exchange Profibus not possible (invalid ad- dress, initialization error) green/red/off...
  • Page 19: Accessories

    Accessories 6 Accessories View the range of accessories for DigiLine on our website. Description Part number Y-Connector M12B to Profibus P 4723 015 Y-Connector with cable M12B to Profibus, 0.5 m P 4723 018 T-Connector M12B to Profibus P 4723 020 Termination resistor M12B for Profibus P 4723 030 Plug and socket M12B for Profibus...
  • Page 20: Technical Data

    Technical data 7 Technical data Parameter CCT 36x PB CCT 37x PB Interfaces RS-485, Profibus DP "Profibus" interface, device side Analog, Binder M12 socket, 5-pin, B-coded Supply: Power consumption max. 15 W Tbl. 9: Technical data CCT gauge Parameter CPT 200 PB PPT 200 PB RPT 200 PB HPT 200 PB...
  • Page 21: Declaration Of Conformity

    Harmonized standards and applied national standards and specifications: DIN EN 61000-6-2:2005 DIN EN 61000-6-3:2007 + A1:2011 DIN EN 61010-1:2010 DIN EN 61326-1:2013, group 1, class B Signature: Pfeiffer Vacuum GmbH Berliner Straße 43 35614 Asslar Germany (Daniel Sälzer) Asslar, 2020-08-28...
  • Page 22: Declaration Of Conformity

    Restriction of the use of certain hazardous substances, delegated directive 2015/863/EU Harmonized standards and applied national standards and specifications: DIN EN 61326-1:2013, group 1, class B DIN EN 50581:2013 Signature: Pfeiffer Vacuum GmbH Berliner Straße 43 35614 Asslar Germany (Daniel Sälzer) Asslar, 2020-08-28...
  • Page 23 23/24...

This manual is also suitable for:

Cct 36x pbCct 37x pb

Table of Contents