Specifications - hatch A50B Operation Manual

Pressure steam sterilizer
Table of Contents

Advertisement

No.:GC-JS-21
2.1 Chamber specification:
Available space size:
Model No.
A50B
2.2 Parameter of Chamber: Design
pressure: -0.1/0.27MPa;
Design temperature:140℃
2.3 Parameter of autoclave:
Max working temperature:137℃
Max working pressure:0.24MPa
Safety valve set-pressure:0.24MPa; Safety valve open-pressure:0.24MPa~0.26MPa
Capacity of clean water tank:1L
Capacity of waste water tank:0.8L
Power input:220V±22V, 50Hz,2900VA
Life span:5 年
2.4 Environmental requirements for use: Environment
temperature: 5℃~40℃;
Relative humidity: ≤85%;
Ambient atmospheric pressure: 80kPa~106kPa;
Water usage requirements as shown in the table below:
Evaporation
Silica
residue
Version: 01/00

2. Specifications

Dimensions(DN * Length mm)
L=175, H=55, D=280
Space available
Iron
Cadmium
Effective date:2020-03-30
Lead
Other
heavy
metals
9