hatch A50B Operation Manual page 26

Pressure steam sterilizer
Table of Contents

Advertisement

No.:GC-JS-21
7.11 Door opening treatment in special cases
If the door cannot be opened due to equipment failure (such as E05 alarm), and the load in the equipment needs to
be taken out, follow the steps below:
To ensure safety, you must first cut off the power and ensure that the pressure in the sterilization chamber is
completely released.
Use a tool to open the safety valve to completely release the air pressure. Be careful not to get burned by the
discharged water vapor. Can be operated with insulated gloves;
1) Power on;
2) Open the door
When opening the door under special circumstances, please ensure that the sterilizer is powered off and
the pressure in the sterilization chamber has been completely released.
7.12 Review or slap of internal records
The device can store 20 recent operation records and print again. When the printing paper is out of paper, you don't
need to worry about using the lack of sterilization records. Just replace the printing paper and follow the steps below
to make up the sterilization records.:
1) Press the "Record" button on "more" pages to enter:
The record name consists of three parts: program name + count + start time, the user can select the record to be
viewed or swat according to the record name.
2) Press the corresponding record name to enter the corresponding record details interface:
7.13 Handling when not in use for a long time
If users need to use the equipment for a long time (greater than or equal to 4 days), please follow the steps below
to handle the equipment:
Version: 01/00
Effective date:2020-03-30
Press the print key to make up this record。
26