hatch A50B Operation Manual page 25

Pressure steam sterilizer
Table of Contents

Advertisement

No.:GC-JS-21
release the pressure to a safe pressure and end the procedure, waiting for the door to open.
7.9 Open the door and unload the loadings
7.9.1 After the program ends, you can turn the door handle counterclockwise to open the door;
7.9.2 Offload
After opening the door after sterilization, the load can be taken out. The operator should use the configured
instrument tray handle or wear insulated gloves or other tools to remove the sterilized items.
The operator should use the configured instrument tray handle or wear insulated gloves or
other tools to remove the sterilized items to avoid burns.
If the sterilized items are taken out immediately after the procedure, there may be a
small amount of condensate on the sterilized items, and the door should be opened and wait for 5
minutes.
7.10 Record reading and storage
7.10.1 Record reading
The record of the printer is the same as the record on the SD card. See the appendix for the reading instructions of
the record:
7.10.2 Record storage
SD card records are recommended to make a backup every 3 months and empty the SD card. Too many records
stored on the SD card may cause the risk of missing records.
The carrier of the records printed by the printer is thermal paper, which should be stored in a dry, normal temperature
environment and avoid direct sunlight. The longest storage life of this record is 2 years.
Version: 01/00
Effective date:2020-03-30
After selecting "Yes", the device will automatically
25

Advertisement

Table of Contents
loading

Table of Contents