hatch A50B Operation Manual page 19

Pressure steam sterilizer
Table of Contents

Advertisement

No.:GC-JS-21
:Used to set the date and time, after entering the page is as below:
6.1.4.2
pressure, the page after entering is like below:
corresponding unit and press OK to change the unit accordingly.
:The user sets whether the power-on preheating function is turned on or off. After entering, the
6.1.4.4
page is like this:
special care should be taken not to touch the sterilization
chamber when the door is open to avoid burns.
Version: 01/00
The left picture shows that the current time is
13:11pm, Mar 30
MM:month DD:day
6.1.4.3
The current unit is indicated in yellow font. Select the
Effective date:2020-03-30
th
2020
YY:year
hh: hour
mm:min
: Setting units for temperature and
When set to "On": The equipment will start to warm
up and keep warm after being powered on, in order
to shorten the sterilization cycle time;
When set to "off": the device will not preheat after
power on, the device will only start to preheat after
the
program
the
entire sterilization cycle time will be 5-7
minutes longer than the "on"
It is recommended to set this setting to "On".
When this setting is turned on,
is
running,
19

Advertisement

Table of Contents
loading

Table of Contents