No.:GC-JS-21
No rolling
Stack cannot exceed 2 layers
Temperature range 5℃~40℃
Relative humidity is no greater than 80%
keep dry
Safety Notices:
This sterilizer is only suitable to sterilize moisture- and heat-resistant medical devices, not okay
for oils and powders, such as petroleum jelly and agar.
This sterilizer cannot be used for sterilization of liquids or liquids sealed with closed containers
(especially glassware), which may cause the containers to burst, endangering the safety of
persons and equipment.
Chloride ion is an important factor causing corrosion of stainless steel. If you use this sterilizer
to sterilize items which contains chloride ions, you must rinse the sterilization chamber with
clean water after each cycle to prevent the deposited chloride ions from corroding the pot and
extending the service life of the equipment.
When you see the symbol anywhere on the device,
is high, avoid touching to avoid burns.
This equipment complies with the emission and immunity requirements of Class A equipment
specified in GB / T 18268. It may cause interference when used in home or similar environment,
and protective measures are required.
It is recommended to evaluate the electromagnetic environment before using the equipment. It
is forbidden to use this device near strong radiation sources (such as unshielded radio frequency
sources), otherwise it may interfere with the normal operation of the device.
Version: 01/00
Effective date:2020-03-30
it means that the surface temperature