Appendix 3:Explanation Of Each Stage - hatch A50B Operation Manual

Pressure steam sterilizer
Table of Contents

Advertisement

No.:GC-JS-21
The time
difference is
Total Cycle time
(18 m : 34 s)
SERIAL
NO.:300012345
OPERATER:
----CLASS B---
COUNT:00001
B.W.
WATER
COUNT
DATE:
01-07-
2020
TIME: 15:46:25
PROGRAM: 134 /
QUICK
Appendix 3:Explanation of each stage
No.
Display
1
Standby
2
Warm up
3
vacuum
4
Temperature rise
5
drain
6
Sterilization
Version: 01/00
HH:MM:SS
START
15:51:43 000 072.6 05:18
VACUUM
15:52:45 -080 079.1 01: 02
RAISE
15:53:00 021 089.2 00:15
VACUUM
15:53:50 -080 095.2 00:50
RAISE
15:54:05 010 103.9 00:14
VACUUM
15:55:09 -080 105.7 01:05
RAISE
15:56:24 221 133.7 01:15 STER-START:
15:56:24 221 133.7 01:15
MAX TEMP:
MIN TEMP:
MAX PRESS:
MIN
PRESS:
207
STER-END
16:00:24
213 134.6 04:00
DRY-END
16:04:58 -017 118.8 04:34
CYCLE COMPLETE
16:04:59
000 119.1 00:01
Explanation
The device is normal and ready for operation
Warm-up phase
Vacuuming
Steam is being injected into the sterilization chamber: temperature and
pressure will increase
The sterilization room is draining
Sterilizing: maintain sterilization temperature and pressure
Effective date:2020-03-30
Pressure Unit
136.1
133.7
During sterilization
227
phase
KPA ℃
Temperature
Unit
Time of vacuum phase
39