Models 59 and 59-T Evacuation Chair
5.2 Track System - Model 59-T Only
The EZ Glide track system enables operators to glide
the chair down stairs instead of lifting and carrying it.
GUIDELINES FOR USING THE TRACK SYSTEM
• Using the chair on stairs requires a minimum of
two trained operators. Ferno recommends using a
third person as a "spotter" (see Gliding the Chair
Down Stairs, page 18).
• Verify that the track system is fully opened and
locked into place before using it.
• Never lubricate track belts. Lubricated track belts
can perform unpredictably, resulting in injury to
the patient and/or operators.
• Moisture, water, snow, ice, or debris on or
between the track and belts can cause irregular
track-belt performance that results in sudden
changes in the weight operators must support.
Make sure the track and track belts are clean and
dry before using the chair on stairs.
• Water, snow, ice, or debris on the stairs can cause
poor footing for operators. To avoid possible
injury, clear the stairs or select an alternate route.
OPENING THE TRACK
1. Grasp the track-release bar located near the
top of the track (Figure 12) and firmly pull it
back until the track locks into the fully extended
position.
2. Verify that the lock is engaged by trying to push
the track closed. If the lock is fully engaged the
track will not move.
CLOSING THE TRACK
With your hand, push down on the red track-closing
handle (Figure 13) until the track closes completely.
© Ferno-Washington, Inc. 234-3298-01 October 2005
Using the Features
Figure 12 - Opening the Track
Figure 13 - Closing the Track
WARNING
!
Using the chair with the track unlocked can
cause injury. Verify that the track is locked
before transporting the patient.
WARNING
!
Lubricating the track system can result in
injury to patient and/or operators. Never
lubricate the track or belts.
13