Waltron 9065C Instruction Manual page 32

Portable luminescent dissolved oxygen analyzer
Table of Contents

Advertisement

Calibration and Compensation
11. In case the O
2
for the O
sensor constants given on the calibration
2
certificate.
a. Use the arrow buttons to enter the digits and the
"Enter" button to move to the next digit.
b. After the last digit, press the "Enter" button to proceed
to the next menu point.
c. Select the menu point "Previous Menu" to return to the
calibration menu.
12. Press the "Enter" button to save the calibration. If the
calibration completes successfully, the 9065C software
program will return to the "main menu", otherwise you
will receive the message: "Questionable calibration, old
calibration restored"
13. You can also leave the "Calibration" menu without storing
the calibration by selecting the "Main menu" option and
pressing the "Enter" button.
31
sensor has been exchanged, fill in the values
O
SENSOR CONSTANTS
2
C1: + 0,86469
C2: - 0,03803
C3: + 0,00852
C4: + 28,98344
PREVIOUS MENU
HOOFDMENU
P-ATM.
ZERO POINT :0.000 %
O
HIGH
2
O2 SENSOR CONSTANTS
STORE CALIBRATION
MAIN MENU
P-ATM.
ZERO POINT :0.000 %
O
HIGH
2
O2 SENSOR CONSTANTS
STORE CALIBRATION
MAIN MENU
9065C Portable Luminescent Dissolved Oxygen Analyzer
CALIBRATION
:1.013 BAR
:4.010 %
CALIBRATION
:1.013 BAR
:4.010 %
Waltron User Manual