Note: When verifying legacy stereo measurements using a blue side stereo tip in a VER2600E block, an orange box will flash around the INDEX if
below minimum set value.
5
– Perform manual length measurements between the crosshairs.
With accurate cursor placement, a measurement tip of any type in good condition should give results of 0.100 ± 0.005 inches (1.00 ± 0.05 mm). If your
measurement does not fall within these limits, try the suggestions under
Troubleshooting Measurements.
Each measurement verification block comes with a calibration schedule and a certificate of calibration (C of C), which notes the calibration date and
due date.
99
Need help?
Do you have a question about the Everest Mentor Visual iQ VideoProbe and is the answer not in the manual?
Questions and answers