Image & Video Setup
1
– Tap the on-screen Logo (or press
2
– Select to change Image and Video related settings and defaults.
3
– Follow the procedure shown here to change either the Image Save Location or Video Save Location.
These represent the two locations where quick-save images or videos are automatically stored. In
the example shown here, these locations are both initially set to the D: drive. Press the Select soft key
and then choose the desired directory path. When the desired storage location is opened, select
the Done soft key to complete the process.
Note: The DVD drive may not serve as either the default or alternative saving location. Files must first
be saved in a directory located in the instrument's memory or on a connected USB storage device.
Files can later be copied and pasted to optical media in an attached DVD drive.
4
– Using the process outlined above, the default Image Save Location is changed to C:\TestImages\
Images\FolderA\. This is the location where quick-saved images are stored
about Saving Images).
5
– Choose either BMP or JPEG as the default image file type.
6
– S e t t h e v i d e o r e c o r d i n g f o r m a t t o e i t h e r H . 2 6 4 H i g h o r H . 2 6 4 L o w .
7
– Determines if the microphone icon appears in the display's bottom right corner
during video recording. When displayed, tap the icon to mute or unmute the recording
of sound along with video.
Click here to learn more about muting
or turning sound ON when recording live video.
8
– When set to Default, the system will normally recognize the
connected Microphone. Alternatively, choose a microphone
from those available (choices are only those devices already
connected to the MViQ or paired via Bluetooth).
9
– When set to Default, the system will normally recognize the connected Speaker. Alternatively,
choose a speaker from those available (choices are only those devices that are already connected
to the MViQ or paired via Bluetooth).
10
– Should a not-yet-defined tip configuration require an alternative Distortion Correction Table,
Contact Technical Support. A table supplied by Technical Support will include instructions on how
to Import the file.
11
– When set to OFF, the menu-directed inspection stage name is not saved in the image but will
still appear during the MDI process.
Executing a White Balance
A white balance corrects the color so that white appears white despite any slight hues that may
exist under the lighting conditions present when performed.
12
– Select New and follow the prompts or select Default to restore factory color settings.