Fluke 7037 User Manual page 21

Calibration bath
Hide thumbs Also See for 7037:
Table of Contents

Advertisement

Press "SET" to select this memory and access the set-point value.
Access set-point value
C
25.00
Press "UP" or "DOWN" to change the set-point value.
Increment display
C
30.00
Press "SET" to accept the new value and display the vernier value. The bath
begins heating or cooling to the new set-point.
Store new set-point, access vernier
0.00000
Press "EXIT" and the bath temperature will be displayed again.
Return to the temperature display
24.73 C
The bath heats or cools until it reaches the new set-point temperature. Set the
heater switch to position "HIGH" to allow the bath to more quickly reach a higher
temperature. The "HIGH" setting may be necessary to reach higher temperatures
and control at high temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cut-out should be correctly set to
prevent this from happening. See the Cutout section.
If operating the bath below 45 °C set the COOLING power switch to ON. The
cooling temperature may require adjustment to provide the proper amount of
cooling. See the Cooling section.
To obtain optimum control stability adjust the proportional band as discussed in
the Proportional Band section.
Current value of set-point 1, 25.00 °C
Current vernier value
Bath temperature display
Calibration Bath
Quick Start
11

Advertisement

Table of Contents
loading

This manual is also suitable for:

703870407041