Fluke 7037 User Manual page 14

Calibration bath
Hide thumbs Also See for 7037:
Table of Contents

Advertisement

7037/7038/7040/7041
User's Guide
ELECTRICAL HAZARD
BATH FLUIDS
4
These guidelines must be followed to ensure that the safety
mechanisms in this instrument will operate properly. This
mains supply for the instrument must be capable of the
power requirements for the instrument as listed in the
Specifications section. The power cord of the instrument is
equipped with a three-pronged grounding plug for your
protection against electrical shock hazards. It must be
plugged directly into a properly grounded three-prong
receptacle. The receptacle must be installed in accordance
with local codes and ordinances. Consult a qualified
electrician.
DO NOT use an extension cord or adapter plug.
DO use a ground fault interrupt device. This instrument
contains a fluid. A ground fault device is advised in case
fluid is present in the electrical system and could cause an
electrical shock
Always replace the power cord with an approved cord of the
correct rating and type. If you have questions, contact an
Authorized Service Center. See the How to Contact Fluke
Calibration section.
High voltage is used in the operation of this equipment.
Severe injury or death may result if personnel fail to observe
the safety precautions. Before working inside the
equipment, turn off the power and disconnect the power
cord.
Fluids used in this bath may produce noxious or toxic
fumes under certain circumstances. Consult the fluid
manufacturer's MSDS (Material Safety Data Sheet). Proper
ventilation and safety precautions must be observed.
The instrument is equipped with a soft cutout (user settable
firmware) and a hard cutout (set at the factory). Check the
flash point, boiling point, or other fluid characteristic
applicable to the circumstances of the bath operation.
Ensure that the soft cutout is adjusted to the fluid
characteristics of the application. As a guideline, the soft
cutout should be set 10 °C to 15 °C below the flash point of
the bath fluid. See the Bath Fluid section, for specific
information on bath fluids and the Cutout section.