Page 16
CONTENTS GENERAL INTRODUCTION………………....16 INITIAL INSPECTION……...………………....16 WATER PROOF…….……...………………....17 INSTALLING THE BATTERIES..…………....17 DISPLAY & KEYS FUNCTIONS…..……………...……18 ..............18 A. Display ..............19 B. Keys OPERATIONAL PROCEDURES..…….......21 ..........21 A. Buffer Set Selection ............21 B. pH Calibration a. Calibration w ith an ATC/Temp probe in the ........21 pH-AUTOLOCK mode b.
GENERAL INTRODUCTION Thank you for selecting the 6010M meter. The 6010M is a precision tool that measure pH, ORP and temperature. A built-in microprocessor stores, calculates and compensates for all parameters related to pH determinations including pH electrode temperature characteristics, electrode slope deviations and buffer solutions.
WATER PROOF Though the 6010M meter is housed in a watertight case, DO NOT use it underwater. The watertight case prevents permanent damage to the unit if accidentally dropped into non-corrosive solutions. Follow these steps immediately if the unit is immersed in any solution: 1.
[Note: Press the “ON/OFF” key to turn the unit on. If the unit is running then you can press the “ ON/OFF ” key to turn the unit off. The unit will automatically turn off after 30 minutes of no key activity.] DISPLAY &...
13. Full- 7. Delete- This will indicate that all 50 To delete stored data. data storage sites are used 14. HOLD- 8. pH/mV- This will indicate that the Unit and mode indicators. reading frozen during AUTOLOCK mode. 15. Recall- 9. Main display for pH, mV and To recall data from the data probe efficiency values storage.
Page 21
STAND/SLOPE- The “STAND” and “SLOPE” keys are used for dual point pH calibration of the unit. Pressing and holding the “STAND” key while turning on the power, will change the buffer set to the other available buffer set. MEA. / EFF.- The key is used to bring the unit out of the AUTOLOCK condition when operating in the pH-AUTOLOCK or mV-AUTOLOCK mode.
OPERATIONAL PROCEDURES A. Buffer Set Selection The 6010M meter has two buffer sets: 7.00, 4.01, 10.01pH and 6.86, 4.00, 9.18pH. The meter is factory pre-set at 7.00, 4.01 and 10.01pH. To change the buffer set, turn off the unit, then press and hold the “STAND”...
exit the calibration mode. Dual point calibration is complete. The unit calculates and compensates for the pH electrode SLOPE deviation corresponding to the values of the two calibration buffers. After calibration, press and hold “MEA./EFF.” key for 5 seconds to display the new electrode efficiency.
Turn the unit on. Press “CLEAR/ENTER” key for 5 seconds, all LCD elements will lit up. The meter clears all calibration values stored in the internal memory. Connect the pH electrode to the BNC connector and the ATC/Temp probe to the ATC/Temp connector of the unit: “ATC”...
is ready to be sloped at the second buffer. Rinse the pH probe in distilled water and immerse it in the second buffer solution (either 4.00/4.01 or 9.18/10.01), then press “SLOPE” key to calibrate. Once the unit calibrates the second point and the unit will automatically exit the calibration mode.
not AUTOLOCK. In this case, use the pH NON- AUTOLOCK mode for measurements.] Measurement with manual temperature compensation in the pH-AUTOLOCK mode. Connect the pH electrode to the BNC connector of the unit. The “MAN” icon will lit up. Set unit to display the sample temperature by pressing the “UP”...
D. Temperature Measurements The 6010M can measure temperature independently with the ATC/temp probe without using the pH electrode. Place the ATC/temp probe in the sample. The unit will display the measured temperature.
Rinse electrode with distilled water and immerse it in sample to be measured. Allow sufficient time for the display to stabilize. The instrument will display the mV value of the sample. F. Save, Recall and Delete Data Saving readings to memory. In pH-AUTOLOCK, mV-AUTOLOCK, pH and mV modes, press the “CLEAR/ENTER”...
pH BUFFERS The temperature coefficient of pH calibration buffers 4.01, 6.86, 7.00, 9.18 and 10.01 are stored inside the instrument. The buffers used to calibrate the instrument must exhibit the same temperature characteristics as the stored values. Temperature coefficient of the pH buffers 4.00 6.86 9.18...
3. Bring sample temperature 3. Measured temperature is into the correct measuring out of the –10.0 to 120.0°C range. range. [Note: If the meter still does not perform normally after the above measures are taken, call Jenco Service Department.]...