JENCO 6179MB Operation Manual

JENCO 6179MB Operation Manual

Ph/orp/ion/ temperature based bluetooth benchtop meter

Advertisement

Quick Links

Operation Manual
MODEL 6179MB
pH/ORP/ION/Temperature
Based Bluetooth Benchtop Meter
6179MB

Advertisement

Table of Contents
loading

Summary of Contents for JENCO 6179MB

  • Page 1 Operation Manual MODEL 6179MB pH/ORP/ION/Temperature Based Bluetooth Benchtop Meter 6179MB...
  • Page 2: Table Of Contents

    CONTENTS GENERAL INTRODUCTION.........2 INITIAL INSPECTION............2 POWER INPUT..............3 INSTALLING THE BATTERIES........3 CONNECTORS..............4 DISPLAY & KEYS FUNCTIONS..........4 ................4 A. Display ........5 B. Operational Keys Description OPERATIONAL PROCEDURES........7 ..........7 A. Buffer Set Selection ............7 B. pH Calibration ...........11 C. pH Measurements ........12 D. Temperature Measurements ...............12 E.
  • Page 3: General Introduction

    GENERAL INTRODUCTION Thank you for selecting the 6179MB meter. The 6179MB is a precision tool that measures pH, mV, ion concentration and temperature. A built-in microprocessor stores, calculates and compensates for all parameters related to pH determinations including pH electrode temperature characteristics, electrode slope deviations, offset and buffer solutions.
  • Page 4: Power Input

    POWER INPUT The model 6179MB can be powered by an UL/CE approved 100 to 240 V AC adaptor as well as 6 “AAA” alkaline batteries. Check the label on the AC adaptor supplied with the instrument to make sure that the AC line voltage is correct.
  • Page 5: Connectors

    CONNECTORS Figure 2: Connector pH/ORP/ISE connector (BNC connector) ATC connector (8 PIN connector) AC adaptor input connector DISPLAY & KEYS FUNCTIONS A. Display 1.68 4.01 7.00 10.01 4.00 6.86 9.18 12.46 Figure 3: Active LCD screen 1. WAIT- 7. AUTO- This will be displayed when AUTOLOCK mode indicator.
  • Page 6: Operational Keys Description

    9. LOCK- 3. pH- This will indicate that the Unit and mode indicators. reading is frozen during AUTOLOCK mode. 10.EFF- 4. mV (RmV)- This will be displayed if the Unit and mode indicators. user is viewing the efficiency of the electrode. 5.
  • Page 7 Mode- Selects display mode. Pressing this key changes display sequentially display pH-AUTOLOCK, (RmV)-AUTOLOCK, ION-AUTOLOCK, pH , mV (RmV), ion, Recall, Delete One and Delete ALL. The calibration values will not be affected by changing the display modes. In the calibration mode, press “Mode” key to exit calibration mode.
  • Page 8: Operational Procedures

    OPERATIONAL PROCEDURES A. pH Buffer Set Selection The 6179MB meter has two buffer sets: 1.68, 7.00, 4.01, 10.01, 12.46 pH and 1.68, 6.86, 4.00, 9.18, 12.46 pH. In pH mode, press “Set/Clear” key to enter the buffer set selection mode. Use “Up” or “Down” key to select the preferred buffer set.
  • Page 9 To continue with two or more point calibration, the first buffer solution has to be either 7.00 or 6.86 pH.] Rinse the pH and ATC/Temp probe in distilled water and immerse them in the second buffer solution (either 4.00 / 4.01 pH or 9.18 / 10.01 pH corresponding to the flashing number on display).
  • Page 10 sloped at the second buffer if the first buffer solution is 6.86 or 7.00 pH. [Note: At this moment, press the “Mode” key, the unit will exit the calibration mode. Single point calibration is complete. To continue with two or more point calibration, the first buffer solution has to be either 7.00 or 6.86 pH.] Repeat steps “e”...
  • Page 11 flashing number on display). Allow temperature reading to stabilize, then press “Slope/Enter” key to calibrate. The unit immediately calibrates the second point, the selected two buffers lit up and the remaining buffer starts to flash. The unit is ready to be sloped at the third buffer. [Note: At this moment, press the “Mode”...
  • Page 12: Ph Measurements

    “Mea./Eff.” key for about 5 seconds to display the new electrode efficiency and offset value. C. pH Measurements To take pH measurements, 6179MB must be calibrated before first use. 1. Measurement with an ATC/Temp probe in the pH- AUTOLOCK mode.
  • Page 13: Temperature Measurements

    Repeat steps “c” and “d” of “Measurement with an ATC/Temp probe in the pH NON- AUTOLOCK mode” D. Temperature Measure The 6179MB can measure temperature independently with the ATC/Temp probe without using the pH electrode. Place the ATC/Temp probe in the sample. The unit will display the measured temperature.
  • Page 14: Mv (Rmv) Measurements

    Connect the ORP electrode to the BNC connector of the unit. Press “Mode” key for the LCD to display “MAN”, “mV” and “AUTO” icons will lit up. The "WAIT" icon begins to flash. Rinse the ORP probe in distilled water and immerse it in the standard solution.
  • Page 15: Ion Calibration

    (RmV) value of the sample. ION Calibration The 6179MB can measure ion concentration when using an ion selective electrode (ISE) for the specific ion of interest. The available ion calibration values are 0.10, 1.0, 10.0, 100.0, and 1000.
  • Page 16 solution will appear on the main display. The “Wait” icon will flash until the meter detects a stable value. When the “Wait” icon disappears and the ion calibration value on the secondary screen moves up to the next value, the first point calibration is complete and the meter is ready for the second point calibration.
  • Page 17: Ion Measurements

    Press “Slope/Enter” key. The mV value of the standard solution. When the mV value of the standard solution is stable, press “Slope/Enter" key to save. The ion calibration value on the secondary screen moves up to the next value. The first point calibration is complete and the meter is ready for the second point calibration.
  • Page 18: Ion Selective Electrodes

    2. Measurement in the ION NON-AUTOLOCK mode. Connect the optional combination ion selective electrode to the BNC connector of the unit. Press “Mode” key until the “MAN” icon lit up. Rinse the ISE with distilled water and immerse it in sample to be measured.
  • Page 19 record. Press “Mode” key to exit “Recall” mode. 3. Deleting data. Press “Mode” key until the “dEL onE” or “dEL ALL” will be displayed. Then press the “Slope/Enter" key to go into the corresponding “Delete one” or “Delete ALL” mode. In the “Delete ALL”...
  • Page 20: Ph Buffers

    B. App Download and Installation Download the companion App for the 6179MB from the Apple Store or Google Play Store. Continue to the next step after the correct App is installed to the tablets.
  • Page 21 1. Prepare the 6179MB for Bluetooth connection. Press the “Power/BLE” key for 2 seconds to turn on the 6179MB device. Press the "Mode" key to go to the measure mode. Press and release the “Power/BLE” Key again to turn on the 6179MB Bluetooth connection. The Bluetooth light on the 6179MB will fast blink and it is ready for Bluetooth connection.
  • Page 22 F. Calibration Screen The Calibration Screen displays past calibration data. From this screen, tap the “Calibrate” button and follow the step-by-step instructions to calibrate the 6179MB. (Fig.2) G. Settings Screen The Settings Screen displays the various settings that affects the measurements of the 6179MB. (Fig.3) Fig.1...
  • Page 23 H. Key Operations 1. Take measurements Tap the “Measure” tab in the navigation bar. 2. Calibrate a. Tap the “Calibrate” tab in the navigation bar. b. Tap the “Calibrate” button. c. Follow the on-screen instructions. 3. Save readings a. Tap the “measure” tab in the navigation bar. b.
  • Page 24 Tap the “Settings” tab in the navigation bar. c. Tap the "Pair" button. d. Wait for the App to discover the 6179MB. Tap the "Connect" button. I. Notes: One meter can be paired to only one App at a time. To put an already-paired meter into pairing mode, one must first close the App it is paired to.
  • Page 25: Fcc Warning Statement

    Bluetooth information. Hence, no pairing is needed the next time when opening the same App. After App is successfully connected to the meter, tablet displays meter reading values on the “Measure” page. On Android OS system, if the connection between the App and the meter is lost for over 2 minutes, force close and reopen the same App for normal operation.
  • Page 26: Error Displays And Troubleshooting

    2000 range. 4. Bring sample temperature into 4. Measured temperature is the correct measuring range. out of the -10.0 to 120.0 °C range. [Note: If the meter still does not perform normally after the above measures are taken, call JENCO representative.]...
  • Page 27: Specifications

    SPECIFICATIONS Display Range Resolution Accuracy -2.00 to 16.00 pH 0.01 pH ±0.01 pH mV (RmV) -1999.9 to 1999.9 mV 0.1 mV ±0.05% F.S. ±0.5% F.S. (mono-valent) 0.01 to 2000 0.01, 0.1, 1 ±1.0% F.S. (di-valent) Temperature -10.0 to 120.0 °C 0.1 °C ±0.2 °C 1.68, 7.00, 4.01, 10.01,12.46...
  • Page 28: Warranty