CONTENTS GENERAL INTRODUCTION INITIAL INSPECTION POWER INPUT INSTALLING THE BATTERIES CONNECTORS DISPLAY & KEY FUNCTIONS A. Display Operational keys description OPERATIONAL PROCEDURES A. Buffer Set Selection B. pH Calibration C. pH Measurements D. Temperature Measurements E. mV offset F. mV (Rmv) Measurements G.
50/60 Hz AC noise rejection. This meter is user-friendly for laboratory application. [ IMPORTANT NOTE: The 6177M has a 10 KΩ thermistor and the 6177EU has a 30 KΩ thermistor. There temperature sensor is not interchangeable.]...
INITIAL INSPECTION Carefully unpack the unit and accessories. Inspect for damages made in shipment. If any damage is found, notify your Jenco representative immediately. All packing materials should be saved until satisfactory operation is confirmed. POWER INPUT The model 6177M/6177EU can be powered by an UL/CE approved 100 to 240 VAC adaptor as well as 6 “AAA”...
1. WAIT- 7. AUTO- This will be displayed when the AUTOLOCK mode indicator. unit is still waiting for a stable reading or end point sensing. 8. Buffer selection- This indicator will flash if the 2. BAT- unit is not yet calibrated. This Low battery indicator.
Page 7
Description Stand & Slope/Enter- These two keys are used for pH and ORP calibration. For their specific functions, please see the "calibration" section in each parameter. S tand In the measure mode, press “Slope/Enter” key for 2 seconds, in order to save reading into the data storage site.
OPERATIONAL PROCEDURES lope A. pH Buffer Set Selection The 6177M/6177EU meter has two buffer sets: 1.68, 7.00, 4.01, 10.01, 12.46 pH and 1.68, 6.86, 4.00, 9.18, 12.46 pH. To change the buffer set, turn off the unit, then press and hold the “Stand”...
Page 9
Press “Mode” key to go to pH-AUTOLOCK mode. “ATC” icon, “pH” icon and “AUTO” icon will lit up. The "CAL" icon begins to flash. Rinse the pH and ATC/Temp probes in distilled water and immerse them in the first buffer solution. Allow temperature reading to stabilize, then press and hold “Stand”...
Page 10
Connect the pH electrode to the BNC connector of the unit. Press “Mode” key to go to pH-AUTOLOCK mode. “MAN” icon, “pH” icon and “AUTO” icon will lit up. The "CAL" icon will begin to flash. Rinse the pH probes in distilled water and immerse it in the first buffer solution.
Page 11
[Note: If the first buffer solution is 7.00 or 6.86 pH, at this moment, press the “Mode” key, the unit will exit the calibration mode. Single point calibration is complete. If the first buffer solution is 1.68, 4.00, 4.01, 9.18, 10.01 or 12.46 pH, at this moment, the unit will automatically exit the calibration mode.
After calibration, press and hold “Mea./Eff.” key for about 5 seconds to display the new electrode efficiency and offset. C. pH Measurements To take pH measurements, 6177M/6177EU must be calibrated before first use. Measurement with an ATC/Temp probe in the pH- AUTOLOCK mode.
Repeat steps 3 to 4 of “Measurement with an ATC/Temp probe in the pH NON- AUTOLOCK mode”. D. Temperature Measure The 6177M/6177EU can measure temperature independently with the ATC/Temp probe without using the pH electrode. Place the ATC/Temp probe in the sample. The unit will display the...
E. mV Offset Turn the unit on. In mV mode, press “Clear” key for 5 seconds, all LCD elements will lit up. The meter clears all calibration values stored in internal memory. Connect the ORP electrode to the BNC connector of the unit, “MAN”...
Press “Mode” key until “mV” or “RmV” and “MAN” icons lit Rinse electrode with distilled water and immerse it in sample to be measured.Remove any air bubbles trapped around the probe by shaking or stirring the probe. Allow sufficient time for the display to stabilize. The instrument will display the mV (RmV) value of the sample.
Press “Mode” key to exit “Delete” mode. H. pH Buffers The temperature coefficient of pH calibration buffers 1.68, 4.00, 4.01, 6.86, 7.00, 9.18 , 10.01 and 12.46 pH are stored inside the instrument. The buffers used to calibrate the instrument must exhibit the same temperature characteristics as the stored values.
1999.9 mV range. 3. Bring sample temperature 3. Measured temperature is into the correct measuring out of the 0.0 to 100.0 °C range. range. [Note: If the meter still does not perform normally after the above measures are taken, call Jenco representative.]...
1 point. Offset ±150 mV Input impedance >1 x 10 Ω Thermistor, 10 kΩ at 25 °C (6177M) Temperature sensor Thermistor, 30 kΩ at 25 °C (6177EU) 6 X 1.5 V AAA Batteries or Power 100 to 240 VAC adapter...