CONTENTS GENERAL INTRODUCTION………………....2 INITIAL INSPECTION……...………………....3 CHARGING THE BATTERY PACK..……....3 BATTERY PACK REPLACEMENT.......4 CONNECTORS..............5 DISPLAY & KEY FUNCTIONS…..……………...…………6 ..............6 A. Display .........8 B. Operational Keys Description OPERATIONAL PROCEDURES..……......10 ............10 A. System setup ............13 B. pH Calibration ..........17 C. pH Measurements ........18 D.
This meter is user-friendly for field, industrial and laboratory applications. [ IMPORTANT NOTE: The 6011B has a 10 KΩ thermistor and the 6011BEU has a 30 KΩ thermistor. Their temperature sensor is not interchangeable.]...
INITIAL INSPECTION Carefully unpack the unit and accessories. Inspect for damages made in shipment. If any damage is found, notify your JENCO representative immediately. All packing materials should be saved until satisfactory operation is confirmed. CHARGING THE BATTERY PACK This meter uses a recharagable battery pack. It is recommended to charge the battery when the "Battery"...
BATTERY PACK REPLACEMENT The life expectancy for 6011B/6011BEU rechargable battery is ≥ 300 charge cycle @ 25 °C. 1. Use a screw driver to remove the four screws and battery compartment cover to expose the battery compartment. (Fig.2) 2. With two fingers, grasp the battery pack connector and pull the connector straight up to disconnect and remove the battery pack.
DISPLAY & KEY FUNCTIONS A. Display Figure 5: Active LCD screen 1. Battery charge. 2. CAL- In the calibration mode. 3. AUTO- 4. RECAL- AUTOLOCK mode indicator. Recalibrate meter. 5. WAIT- 6. √- This will be displayed when This will indicate that the the unit is still waiting for a reading is frozen during stable reading or end point...
Page 8
8. DAY- 7. %- Represents the time unit Efficiency of the electrode "days" required for unit. recalibration. 10. MAN- Manual temperature Data storage site number. compensation. 11.ATC- Auto temperature 12. Buffer sets. compensation. 13.Delete- 14. TIME- In the delete mode. In the time setting mode.
Operational Keys Description Figure 6 Description Power/BLE(Bluetooth)- Press and hold this key for 3 seconds to power on and shut off the meter. In the measure mode, press this key to turn Bluetooth on or off. Mea./Eff.- This key is used to bring the unit out of the AUTO condition when...
Page 10
Mode- This key is used to selects display mode. Pressing this changes display sequentially to display pH-AUTOLOCK, mV (REL mV)-AUTOLOCK, pH, mV (REL mV). In the “Recall” and “Delete” modes, press this key to exit “Recall” and “Delete” modes respectively. In the calibration mode, press “Mode”...
OPERATIONAL PROCEDURES A. System Setup Use the System Setup to customize operation of the 6011B/6011BEU meter. Press the "SET" key to enter system setup mode and enter first setup, RECAL. 1. RECAL setup a.
Page 12
selected by user. d. “Year” will flash first. Press "Up" or "Down" key to set the correct year. Then press "Enter" key to save. e. “Month” will flash. Press "Up" or "Down” key to set the correct month. Then press “Enter” to save. “DAY”...
Page 13
corner will show the 5 selected buffer set. [Note: 5 buffer tubes display is depended on selected buffer set.] Temperature unit setup This function allows end user to select the between "°C" or "°F" temperature unit. “SET” icon will lit up. Main LCD will show “Atc”. Lower LCD will show “unit”...
Temperature unit : °C Clear all CAL and user records B. pH Calibration The 6011B/6011BEU uses up to 5 point calibration. [Note: If the device is required to perform more than one calibration points, the first calibration point must be 6.86/7.00 pH.]...
Page 15
flashing number on display). Allow temperature reading to stabilize. Then press “Enter” key to calibrate. The “WAIT” icon will flash until the unit detects a stable reading. Once the unit calibrates the second point. The selected two buffers lit up and the remaining buffer starts to flash. The unit is ready to be sloped at the third buffer.
Page 16
To continue with two or more point calibration, the first buffer solution has to be either 7.00 or 6.86 pH.] Repeat steps “e” of “Calibration with an ATC/Temp probe in the pH-AUTOLOCK mode” for 2 to 5 point calibration. The unit calculates and compensates for the pH electrode slope deviation corresponding to the values of the calibrated buffers.
Page 17
[Note: At this moment, press the “Mode” key, the unit will exit the calibration mode. Dual point calibration is complete.] The third point, the fourth point and the fifth point are the same as the second point calibration. When the unit completed the fifth point calibration, press the “Mode”...
“Mea./Eff.” key for about 5 seconds to display the new electrode efficiency and offset value. C. pH Measurements To take pH measurements, 6011B/6011BEU must be calibrated before first use. 1. Measurement with an ATC/Temp probe in the pH AUTOLOCK mode.
Repeat steps “c” to “d” of “Measurement with an ATC/Temp probe in the pH NON- AUTOLOCK mode”. D. Temperature Measurements The 6011B/6011BEU can measure temperature independently with the ATC/Temp probe without using the pH electrode. Place the ATC/Temp probe in the sample. The unit will display the measured temperature.
Connect the ORP electrode to the BNC connector of the unit. Press “Mode” key for the LCD to display “MAN” and “mV” icons will lit up. The "WAIT" icon begins to flash. Rinse the ORP probe in distilled water and immerse it in the standard solution.
Press “Mode” key until the “mV (REL mV)” and “MAN” icons lit up. Rinse the ORP electrode with distilled water and immerse it in sample to be measured. Remove any air bubbles trapped around the electrode by shaking or stirring the electrode.
2. Locating data files in Windows Explorer Navigate to Windows Explorer and the PC will recognize the meter as the “Jenco” removable drive. There is one CSV file that can be copied to a location on your PC: Record contains all of the data currently stored on the meter, including the most recent 5 calibration records for pH and the most recent 5 calibration records for mV.
App when connected to tablet or mobile phone. B. App Download and Installation Download the companion App for the 6011B/6011BEU from the Apple Store or Google Play Store. Continue to the next...
Page 24
“Power/BLE” Key again to turn on the 6011B/6011BEU Bluetooth connection. Bluetooth icon 6011B/6011BEU will fast blink and it is ready for Bluetooth connection. 2. Pair the 6011B/6011BEU to the Tablet or Mobile phone. Tap “Settings” tab at in the navigation bar of the App screen.
Page 25
D. Key Operations 1. Upload and save data Tap the “Saved” tab in the navigation bar. "Sync Data" button is displayed. Tap "Sync Data" button. The App will display the loading diagram and the estimated time at the bottom. The display shall return to the Saved screen when the data transmission is completed.
Page 26
E. Notes: One meter can be paired to only one App at a time. To put an already - paired meter into pairing mode, one must first close the App it is paired to. Turn on tablet or mobile phone and activate Bluetooth. (Select Optimized Power Mode on Android tablets or mobile phone).
FCC WARNING STATEMENT Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment NOTE: THE MANUFACTURER IS NOT RESPONSIBLE FOR ANY RADIO OR TV INTERFERENCE CAUSED UNAUTHORIZED MODIFICATIONS TO THIS EQUIPMENT. SUCH MODIFICATIONS COULD VOID THE...
/ replace electrode.] Change the battery pack. The total usage time is significantally shorter on a fully charged battery pack. [Note: If the meter still does not perform normally after the above measures are taken, call JENCO Service Department.]...