4.6

How to Shut Down the System

1.
2.
3.
Note
Caution:
Caution:
MAN-03270 Revision 003
Select File>Exit.
Select Power off the computer? from the dialog box. (See the figure Exit from
Acquisition Station Dialog Box on page 40.
Select Yes.
Chapter 4—Startup, Functional Tests, and Shutdown
If a message appears and indicates that there are spool jobs,
select log off or wait until the jobs are finished. (The spool
continues again later.) Hologic recommends that you allow
the jobs to complete before shut down.
Do not use a circuit breaker or Emergency Off switch as a
routine method to turn off the Selenia.
Never turn off the Acquisition Workstation or Value
Console Circuit Breaker except in emergency. The circuit
breaker can turn off the Uninterruptible Power Supply
(UPS) and risk data loss.