InSight FD Mini C-arm User Guide
Using the System
11.4 Powering Up the System
Warning:
The system should be at ambient room temperature.
The power cord must be plugged into a standard single-phase three-wire
grounded outlet. A special dedicated outlet is not required.
To power up the system, press the Computer On/Standby switch.
InSight FD will prompt the user to log in.
If the system has been off for more than two weeks, use the following
source conditioning procedure to safeguard and prolong the life of the
X-ray source:
Table 6
Conditioning Schedule
Step
1
2
3
4
5
Caution
11.5 Powering Down the System
To close the application and place the system in standby:
• From the InSight FD Main Screen click the Shut Down button.
• From the Exit Menu on the InSight FD Main Screen, click Exit
22
The InSight FD system operates on standard 115 Volt, 60 Hz AC
power for United States, Japanese and Canadian models, 230 Volt,
60 Hz AC, and 240 volt, 50 Hz AC power for other export models.
• The system should only be plugged into a properly
grounded three-wire wall outlet.
• Never cut the third (grounding) pin from the power
cord to plug the unit into an ungrounded two wire
outlet. Never use a three wire-to-two wire adapter.
kV
mA
40
0.05
50
0.10
60
0.10
70
0.10
75
0.10
The system must warm up for FOUR MINUTES before
generating X-rays. Failure to provide adequate warm up could
result in damage to the system.
before unplugging the power cord.
Minimum Time
2 minutes
2 minutes
2 minutes
3 minutes
3 minutes
MAN-04858 Revision 006