Setup; Adjustment - Version X, C - Manual Switchover - Vega VEGASOURCE 81 Operating Instructions Manual

Source holder
Table of Contents

Advertisement

Switching the radiation
on
VEGASOURCE 81 • Source holder
6

Setup

6.1
Adjustment - Version X, C - Manual
switchover
Warning:
Before switching on the radiation, make sure that no persons are
inside the radiated areas (also not inside the vessel).
Radiation must only be switched on by trained personnel.
The figures in brackets refer to the following illustration.
1
3
Fig. 24: Switching the radiation on - VEGASOURCE 81 version X, C
1
Loosen safety screw
2
Fold up the safety screw
3
Turn stirrup by 180°
4
Fold up and tighten safety screw.
1. Initial situation: Source holder is in position OFF
2. Open and remove padlock
Note:
The combination lock can be opened with the number combination
000 on delivery.
We recommend changing the number combination to a personal
number sequence. Make a note of the number combination and keep
it in a safe place.
3. Loosen safety screw (1)
4. Fold up the safety screw (2)
5. Turn the stirrup clockwise by 180° (3)
2
4
6 Setup
33

Advertisement

Table of Contents
loading

Table of Contents