Connect the oscilloscope to the measuring points of the (CPU) D916 TP "KV" and
"KVS" (also see the troubleshooting instructions). Oscilloscope setting: Channel 1 = TP
KV 1V/div, channel 2 TP KVC 2V/div, trigger channel 2, trigger stage 2.5 V, 2V/div, 50
ms/div (oscilloscope setting: Channel 1 = TP KV 1V/div, channel 2 TP KVC 2V/div, trig-
ger channel 2, trigger stage 2.5 V, 2V/div, 50 ms/div)
Follow the same procedure as described in "kV meter method".
3.7.2

mAs accuracy

Requirement:
The measured mAs values must be within the limit values, see tables.
One of two measuring methods can be used:
1.) mAs meter method:
- mAs meter (e.g., MAS meter 8160400 with a measuring accuracy of 1 % ± 1
decimal place)
2.) Oscilloscope method:
- Storage oscilloscope with a measuring accuracy of ± 2.5%
Oscilloscope method:
Open the top cover of the MOBILETT XP Digital and place it sideways on a table or the
like. Do not disconnect any cables.
Connect the oscilloscope to the "JR" and "GND" measuring points on CPU D916.
Switch on the MOBILETT XP Digital.
Select the analog operating mode.
X
Release exposures according to the table.
Parameter setting
MOBILETT XP Digital
Oscilloscope setting: Channel 1 = TP "KV" 1V/div, channel 2 TP "KVS" 2V/div,
trigger channel 2, trigger stage 2.5 V, 50 ms/div
After switching the unit on, wait approx. 5 minutes for it to stabilize.
kV
40
81
133
SPR8-230.831.30.03.02
11.05
Inspection and maintenance
Allowable mAs value
mAs
5
2
10
Page 62 of 74
CS SD 24
4.7 - 5.3
1.9 - 2.1
9.5 - 10.5