Required Tools, Test Equipment And Aids - Siemens MOBILETT XP Digital Maintenance Instruction

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

General maintenance information
2.2

Required tools, test equipment and aids

Standard service tool kit
Measuring device for leakage current and patient leak-
age current, e.g.:
Ground wire tester or e.g.:
2.1 mm precision copper filter (or alternatively, copper
filter set 4406 120 RV090)
Dose meter, e.g.: PTW DIADOS
Densitometer, e.g., DensiX-LE 52003
Digital multimeter, e.g.: Fluke 8060 A
Centering cross
mAs meter (e.g.: 81 60 400) or 2-channel storage oscil-
loscope with ± -2.5% accuracy
kV meter (works with the filter comparison method) or
2-channel storage oscilloscope with +/- 2.5% accuracy
2-channel storage oscilloscope with 2.5% accuracy,
e.g.: Fluke Scope Meter 199 CM
Spring balance for 350 N
Torque wrench for 8-40 Nm, e.g.:
Calibration tool for the "DAP measurement system" op-
tion
Adjuster for tensioning the drive belt
Rope (min. length 4 m; min. tensile strength 1000 N)
Loctite 242
Viscogen oil
2 pieces of wood, approx. 50 x 75 x 500 mm
Siemens AG
Medical Solutions
Lux meter (USA only)
SPR8-230.831.30.03.02
11.05
51 38 727 Y0766
51 38 727 Y0766
90 00 598 RV090
97 17 612 Y0388
49 51 286 Y0388
97 02 101 Y4290
96 60 051 RE999
73 92 074
99 00 846 RE999
65 84 978
65 64 301
MOBILETT XP Digital
Page 17 of 74