Inspection and maintenance
3.3
Function inspection
3.3.1
Operating data
Activating service programs
•
Switch off the system, unplug the power plug, remove the top cover and place it side-
ways on a table or the like. Do not disconnect any cables.
•
Move service switch SW2B into position 2 (service) board D916 (CPU)
•
- Switch the system on.
•
Use the mAs+/mAs- buttons to select the desired service program.
NOTE
Fig. 16: CPU D916
3.3.1.1
Reading out the operating information
PMF
Exposure counter
•
Select ServiceProgram 02 and record the "exposure" counter status in the protocol.
PMF
Error memory
•
Select ServiceProgram 03 and read out the error memory. If a defective component
message is displayed, check it according to the troubleshooting instructions. Reset the
memory in service program 04.
•
Switch off the system.
•
Return service switch SW2B D916 (CPU) to position 1 (user).
Siemens AG
Medical Solutions
"Pr 1" is displayed on the control panel.
The X-ray function is deactivated during an active service pro-
gram.
For detailed information, see Troubleshooting Instructions
SPR8-230.840...
SPR8-230.831.30.03.02
11.05
MOBILETT XP Digital
Page 49 of 74
CS SD 24
49
(Fig. 16 / p.
49).