babyPAC
basic operations
B

BASIC SETTINGS |

- 1 second. Expiration time - 2 seconds. Inspiration Pressure Control - 20 x100Pa (20 cmH
Control - 30 x100Pa (30 cmH2O). O2 - 50% using the white scale.
A

AIR AND OXYGEN |

is attached the Oxygen supply symbol turns to white.
S

SYSTEM CHECK |

of the high priority alarm should sound and the orange silenced indicator should flash for 60 seconds. Check that flow is
coming from the end of the patient circuit.

INSPIRATORY HIGH PRESSURE CHECK |

I
reading between 15-25 cmH
move inspiration pressure control to 40, observe a reading between 25-35 cmH
will sound and the red high pressure visual alarm will sound and flash. Return the inspiration pressure control to 20.
C

CYCLE CHECK |

Please check at this point that the babyPAC
Expiration dials to their extreme, check that the ventilator has responded and no irregularities occur.
O

OPEN CIRCUIT /LOW PRESSURE CHECK |

with no obstruction at the patient circuit end. After approximately 8 seconds the Low Pressure (disconnection) Alarm sounds.
P

PEEP/CPAP CHECK |

occlude the patient circuit. The manometer should read between 6-14. Set the function dial to IMV+CPAP and the expiratory
time to 1 second, check that the expiratory time is approximately every 10 seconds.
S
SET YOUR babyPAC
B100
These are positioned as follows: Function selector - 0. PEEP/CPAP Control - 0. Inspiration time
If Air (black hose) is attached the air supply symbol turns to black and white check. If Oxygen (white hose)
Move function selection switch to 'CMV+ACTIVE PEEP'. All the alarm lights flash in turn, a single burst
Inspiratory pressure check - Keep the patient circuit occluded and observe a
O on the manometer. NO ALARM SHOULD SOUND! Leave the high pressure alarm control at 30,
2
After running the babyPAC for at least 60 seconds, allow the babyPAC
Occlude the patient circuit - Check PEEP - turn the PEEP/CPAP control clockwise to the first click then
AS NOTED IN THE BASIC SETTINGS ABOVE.
O on the manometer. The pneumatic alarm
2
regularly operates about every 3 seconds. Turn the Inspiration/
O).
High Pressure Alarm
2
to cycle

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Pneupac babyPAC B100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Smiths Medical Pneupac babyPAC B100

  • Page 1 babyPAC B100 ™ basic operations BASIC SETTINGS | These are positioned as follows: Function selector - 0. PEEP/CPAP Control - 0. Inspiration time - 1 second. Expiration time - 2 seconds. Inspiration Pressure Control - 20 x100Pa (20 cmH High Pressure Alarm ...
  • Page 2 Prior to the use of this equipment, ensure that the relevant user/installation manuals are read and fully understood. Smiths Medical International Ltd. Find your local Smiths Medical is part of the global technology business Smiths Group plc. Product(s) described may not be licensed or available for sale in Canada or other countries contact information at: Ashford, UK TN25 4BF outside of the United States.

Table of Contents