Basic diagnostic x-ray system table, powered (32 pages)
Summary of Contents for Protec 7073-9-000 Series
Page 1
Diagnostic X-ray system mechanics Model/ID: 7073-9-000X Basis-UDI-DI: 426050264X013ZJ Instructions for use Ident. No. 5073-0-0002 Version: 5.0 PROTEC GmbH & Co. KG Issued: 2021-11-24 In den Dorfwiesen 14, 71720 Oberstenfeld, Germany Subject to alterations Telephone: +49 (0) 7062/92 55-0 E-Mail: protec@protec-med.com...
Page 2
All sheets of this document contain proprietary and confidential information of PROTEC GmbH & Co. KG and is intended for exclusive use by current PROTEC GmbH & Co. KG customers. Copying, disclosure to other or other use is prohibited without the express written authorization of PROTEC´s law department.
Adjustment of the focus-film distance (SID) ....................16 4.2.3.5 Adjusting the light-/beam field.......................... 16 4.2.3.6 Exposure preparation / Exposure releasing ....................17 4.2.3.7 Exposure with cassette on the table top ...................... 17 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 3 von 31...
Page 4
Description of symbols, labels and abbreviations ............29 Symbols ....................................... 29 Type label ......................................30 Labels ........................................30 Position of the signs and labels ............................31 Abbreviations ....................................31 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 4 von 31...
PROTEC GmbH & Co. KG. Assembly personnel and other persons who are not employees of the technical service department of PROTEC GmbH & Co. KG are requested to contact the local branch of PROTEC GmbH & Co. KG before assembly or service work is started.
Hardware and Network System Requirements As a stand-alone product, the PROGNOST C has no hardware or network connection and therefore no hardware or network requirements. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 7 von 31...
1.2.3 Installation NOTE The installation of the PROGNOST C must be performed by PROTEC service department or a service company authorized by them. For more information, please see separate Installation manual of the PROGNOST C. Contact information of persons qualified to perform installations are available upon request at: PROTEC GmbH &...
Appropriate users can be, for example: X-ray technicians, X-ray assistants, medical technical X-ray assistants, surgeons, casualty surgeons, orthopaedists and other trained medical personnel. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 9 von 31...
Settings and calibrations that are not described in these instructions for use must be carried out using the technical description of the device by PROTEC service department or a service company authorized by them.
X-ray system, the connected X-ray generator or the Power Box must be switched off. 2.1.2 Device Operation In case of a malfunction, do not use the PROGNOST C anymore and notify PROTEC service department or a service company authorized by them. 2.1.2.1 Operating Type The PROGNOST C is not intended for continuous operation.
The PROGNOST C is intended for use in an environment in professional health care facilities (e.g., clinics, surgery centres, physiology practices ...). PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 13 von 31...
Brake for the tube head rotation around the carrying axis Free movement (sensor beneath the tube head) Permits the horizontal movement of the tube column and vertical movement of the tube head PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 14 von 31...
Vertical Bucky Wall Stand Brake for the vertical movement of the wall stand Bucky. Up/down release button PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 15 von 31...
Use the adjusting knobs to set the collimator shutters to the size of the cassette being used. The setting is made on the scale for the corresponding focus-film distance (SID). This limits the light- /beam field to the cassette size used. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 16 von 31...
The PRS 500 C is switched on via the control panel of the generator. All system components are supplied with voltage via the generator. If a power box is included in the system, the power is supplied via the power box. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 17 von 31...
Page 18
After successful completion of the self-test, the parameters are displayed NOTE The mechanics for basic diagnostic X-ray systems PROGNOST C must be switched PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 18 von 31...
(e.g., torn cable, bent parts) or if the resulting image quality (e.g., artifacts in the image) is insufficient. In this case, please contact PROTEC service department or a service company authorized by them immediately.
Technical Description in chapter 3 Maintenance and Safety Inspection. In the event that the planned maintenance is not carried out, PROTEC GmbH & Co. KG assumes no liability for damage to the user or third parties, if damage results from inadequate or not carried out maintenance.
The PROGNOST C is designed for a service life of 7 years when used in accordance with the specifications and regular maintenance by PROTEC service department or a service company authorized by them. After the product has reached the end of its service life, further use is at your own risk.
The device is suitable for use in facilities other than Class A residential areas and those that are directly connected CISPR 11 to the public supply network, which also supplies Harmonic emissions Class A PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 22 von 31...
Page 23
150 kHz bis 80 EN 61000-4-6 3 V/m See table below Radiated RF 1kHz 80% AM disturbances 3 V/m EN 61000-4-3 80 MHz to 2.7 GHz PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 23 von 31...
Page 24
2450 2400 – 2570 b/g/n, 217 Hz RFID 2450, LTE Band 7 5240 Pulse modulation: 5500 5100 - 5800 WLAN, 802.11 a/n 217 Hz 5785 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 24 von 31...
V5.0_2021-11-24 PROGNOST C Instructions for use 5073-0-0002 Technical Data Dimensions Room plan version 1 (operation of the system): PATIENT PATIENT PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 25 von 31...
Page 26
To mount or remove the table top, the room must either have a size of at least 6000mm or it is necessary to move components to have more space available. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 26 von 31...
± 135° Tube stand longitudinal travel: 1300 mm Vertical X-ray system image receptor stand Column height: 1850 mm Vertical shift film centre: 400mm - 1735mm PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 27 von 31...
Environmental conditions for shipping and storage Ambient Temperature - 10°C to + 70°C Relative humidity 10% to 95% (non-condensing) Atmospheric pressure 500 hPa to 1060hPa PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 28 von 31...
Caution: pinch-/crushing hazard for hands and fingers Do not exceed the maximum indicated weight Do not exceed the maximum indicated weight Manufacturer Medical Device Order reference Serial number Unique Device Identification PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 29 von 31...
Label on the table top and tube support, floor stand: Caution: Watch out for possible crushing hazards to fingers or hands while moving the table top, table or X-ray unit. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 30 von 31...
Position of the signs and labels Type label Abbreviations Millimetre Centimetre Pound Kilogram °C Degree -Celsius Hectopascal German Industry Standard European Standard CE-Mark Hertz Duty cycle Ampere Serial number PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 31 von 31...
Need help?
Do you have a question about the 7073-9-000 Series and is the answer not in the manual?
Questions and answers