Protec PROGNOST E 7043-5-87 Series Instructions For Use Manual

Protec PROGNOST E 7043-5-87 Series Instructions For Use Manual

Basic diagnostic x-ray system table, powered
Table of Contents

Advertisement

Quick Links

PROGNOST E
Basic diagnostic X-ray system table,
powered
Model/ID: 7043-5-87xx
Basic UDI-DI: 426050264X010ZC
Instructions for use
Ident. Nr. 5043-0-8002
Version: 8.0
PROTEC GmbH & Co. KG
Issued: 2024-04-08
In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
Subject to alterations
Telephone: +49 (0) 7062/92 55-0 E-Mail: protec@protec-med.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PROGNOST E 7043-5-87 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Protec PROGNOST E 7043-5-87 Series

  • Page 1 Basic diagnostic X-ray system table, powered Model/ID: 7043-5-87xx Basic UDI-DI: 426050264X010ZC Instructions for use Ident. Nr. 5043-0-8002 Version: 8.0 PROTEC GmbH & Co. KG Issued: 2024-04-08 In den Dorfwiesen 14, 71720 Oberstenfeld, Germany Subject to alterations Telephone: +49 (0) 7062/92 55-0 E-Mail: protec@protec-med.com...
  • Page 2 All sheets of this document contain proprietary and confidential information of PROTEC GmbH & Co. KG and is intended for exclusive use by current PROTEC GmbH & Co. KG customers. Copying, disclosure to others or other use is prohibited without the express written consent of PROTEC´s law department.
  • Page 3: Table Of Contents

    Operation at the PROGNOST E..........................19 Function of the PROGNOST E..............................20 4.3.1 Switching the PROGNOST E on and off ....................... 20 Safety and Maintenance ....................21 Introduction ...................................... 21 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 3 von 32...
  • Page 4 Description of Symbols, Labels and Abbreviations ............. 29 Symbols ....................................... 29 Type Label ......................................30 Labels ........................................30 Positions of the Signs and Labels............................31 Abbreviations ....................................32 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 4 von 32...
  • Page 5: Revision Status

    Picture / labeling changed due to new handles, Chap. 3.3 Chapter renamed. Chap. 1.2.1 Optional Accessories corrected 2022-12-07 Positioning of the tabletop 2024-04-08 Chap 4.2.1 optimised PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 5 von 32...
  • Page 6: General Notes

    PROTEC GmbH & Co. KG. Assembly personnel and other persons who are not employees of the technical service department of PROTEC GmbH & Co. KG are requested to contact the local branch of PROTEC GmbH & Co. KG before assembly or service work is started.
  • Page 7: Device Description

    Hardware and Network System Requirements As a stand-alone product, the PROGNOST E has no hardware or network connection and therefore no hardware or network requirements PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 7 von 32...
  • Page 8: Installation

    1.2.3 Installation NOTE The installation of the PROGNOST E must be performed by PROTEC service department or a service company authorized by them. For more information, please see separate “Installation manual PROGNOST E.” Contact information of persons qualified to perform installations are available upon request at: PROTEC GmbH &...
  • Page 9: Clinical Benefit

    In den Dorfwiesen 14 | 71720 Oberstenfeld Germany Phone: (+ 49) 7062 – 92 55 0 Fax: (+ 49) 7062 – 92 55 60 E-Mail: protec@protec-med.com Internet: www.protec-med.com PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 9 von 32...
  • Page 10: Safety Instructions

    Settings and calibrations that are not described in these instructions for use must be carried out using the technical description of the device by PROTEC service department or a service company authorized by them.
  • Page 11: General Safety Instructions

    X-ray system, the connected X-ray generator or the Power Box must be switched off. 2.1.2 Device Operation In case of a malfunction, do not use the PROGNOST E anymore and notify PROTEC service department or a service company authorized by them. 2.1.2.1 Operating Type The PROGNOST E is not intended for continuous operation.
  • Page 12: Electromagnetic Environment And Influencing Of Devices

    The PROGNOST E is intended for use in an environment in professional health care facilities (e.g. clinics, surgery centres, physiology practices ...). PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 12 von 32...
  • Page 13: Control Elements And Displays

    2 Signal LED 3 Emergency stop switch 4 Tabletop 5 Foot switch 6 Bucky cassette tray 7 Short hand grip (optional) 8 Bumper profile (optional) PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 13 von 32...
  • Page 14: Foot Switch

    Only by switching off the power supply of the generator, the table safely disconnected from the power supply. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 14 von 32...
  • Page 15 Allow the device to cool down at standstill until Duty cycle of the lifting columns signal LED changes from red exceeded. flashing rhythm to bright green again. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 15 von 32...
  • Page 16: Acoustic Status Indications

    NOTE State message 2,4 and 5: If these state messages can be deleted with the foot switch button and then occur again, the PROTEC authorized service must be notified and the table must be taken out of order. 3.3.3.1 Acoustic Status Indications For all signal lamps state messages, a one-time acoustic message is given by the built-in beeper.
  • Page 17 (4) or by setting the cassette positioner to the corresponding cassette size. The movement range of the Bucky is 545 mm. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 17 von 32...
  • Page 18: Handling

    If the X-ray system table must be operated in a seated position, it is essential to ensure that the tabletop is positioned at the rear. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 18 von 32...
  • Page 19: Zero Balance Performance With The Foot Switch

    Slide the Bucky cassette drawer to the other side. o Position the tabletop as centred as possible (back/front). o The patient should ascend and descend in the middle of the tabletop (green arrow). PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 19 von 32...
  • Page 20: Function Of The Prognost E

    Function of the PROGNOST E 4.3.1 Switching the PROGNOST E on and off The PROGNOST E starts with applying of a power supply and is not started separately. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 20 von 32...
  • Page 21: Safety And Maintenance

    (e.g., bent parts) or if the resulting image quality (e.g., artifacts in the image) is insufficient. In this case, please contact PROTEC service department or a service company authorized by them immediately. Cleaning and Disinfection...
  • Page 22: Inspection And Maintenance

    12 Months by the PROTEC service department or a PROTEC authorized service provider. In the case that the intended safety-related checks are not carried out, PROTEC GmbH & Co. KG accepts no liability for damage to the operator and third parties if and insofar as damage results from insufficient or non-performed checks.
  • Page 23: Maintenance

    Technical Description in chapter 3 Maintenance and Safety Inspection. In the event that the planned maintenance is not carried out, PROTEC GmbH & Co. KG assumes no liability whatsoever for damage to the user and third parties if damage results from inadequate or not carried out maintenance.
  • Page 24: Power Supply

    CISPR 11. This device can cause radio interference in EN 61000-3-3 residential areas, in which case it must be necessary to take appropriate remedial measures such as new PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 24 von 32...
  • Page 25 80 MHz to 2.7 GHz NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structure, objects and people. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 25 von 32...
  • Page 26 2450 2400 – 2570 b/g/n, 217 Hz RFID 2450, LTE Band 7 5240 Pulse modulation: 5500 5100 - 5800 WLAN, 802.11 a/n 217 Hz 5785 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 26 von 32...
  • Page 27: Technical Data

    Longitudinal displacement of the tabletop (from the centre position): ± 470 mm (226 cm tabletop) Longitudinal displacement of the tabletop (from the centre position): The brakes of the tabletop are actuated electromechanically. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 27 von 32...
  • Page 28: Attenuation Equivalent

    Environmental Conditions for Shipping and Storage Ambient Temperature - 10°C to + 70°C Relative humidity 10% to 95% (non-condensing) Atmospheric pressure 500 hPa to 1060hPa PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 28 von 32...
  • Page 29: Description Of Symbols, Labels And Abbreviations

    Order reference Serial number Unique Device Identification Production date Classification according to EN 60601-1 (type B applied part) Caution: pinch-/crushing hazard for hands and fingers PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 29 von 32...
  • Page 30: Type Label

    Release tabletop brakes Type Label Labels Labels on the front side of the different tabletops Carbon fibre tabletop Composite fibre tabletop 200cm Composite fibre tabletop 226cm PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 30 von 32...
  • Page 31: Positions Of The Signs And Labels

    (Composite fibre tabletop). Maximum allowed patient weight (distributed load) for the tabletop (Carbon tabletop). Labels on the front plate Company label Positions of the Signs and Labels PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 31 von 32...
  • Page 32: Abbreviations

    Hectopascal German Industry Standard European Standard CE-Mark Hertz Duty cycle Ampere Serial number V AC Volt (alternating current) V DC Volt (direct current) inch Inches PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 32 von 32...

Table of Contents