Download Print this page

Yamato CPC-2000 Technical Manual

Hide thumbs Also See for CPC-2000:

Advertisement

Quick Links

CPC-2000
www.yamatocorp.com
YAMATO CORPORATION
1775 S. Murray Blvd.
Colorado Springs, CO 80916 USA
Tel (719) 591-1500 Fax (719) 591-1045
Technical Manual
YAMATO TECH CORPORATION
#112-19425 Langley By-Pass
Surrey, B.C. V3S 6K1
Tel (604) 533-2338 Fax (604) 533-0827
Canada
YCO-1094-04-02

Advertisement

loading
Need help?

Need help?

Do you have a question about the CPC-2000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Yamato CPC-2000

  • Page 1 CPC-2000 Technical Manual YAMATO CORPORATION YAMATO TECH CORPORATION 1775 S. Murray Blvd. #112-19425 Langley By-Pass Colorado Springs, CO 80916 USA Surrey, B.C. V3S 6K1 Canada www.yamatocorp.com Tel (719) 591-1500 Fax (719) 591-1045 Tel (604) 533-2338 Fax (604) 533-0827 YCO-1094-04-02...
  • Page 2: Table Of Contents

    CPC-2000 Technical Manual Table of Contents Introduction Setup Technician Setup Mode Setting the Keywords Calibrating the Scale III. Troubleshooting IV. Wiring Diagram Parts List SAFETY INSTRUCTIONS Before using the scale, carefully read, understand, and follow the “Safety Instructions” described in this manual. Observe the advice given in the Operation Manual to ensure proper operation.
  • Page 3: Introduction

    I Introduction The CPC-2000 is available in a 4 lb (2 kg) model and a 10 lb (4 kg) model. It displays units of ki lo - grams, pounds and ounc es (not legal for trade), decimal pounds, and con tin u ous ounces. Any of these units can be enabled or disabled in the technician set-up mode.
  • Page 4 CPC-2000 Technical Manual II Setting the Scale Functions Technician Setup Mode From the front, rotate the upper housing clockwise (to the right) until it rests on the table. Be careful not to stress the wires that connect the main circuit board (in the upper housing) to the power board and load cell (in the lower housing).
  • Page 5: Setting The Keywords

    If “OFF” is selected, the auto-off function will be disabled. The “tran” and “t.rate” keywords do not function on the CPC-2000. The “CAP” setting selects the capacity, in kilograms, of the load cell installed in the scale. Only the 2 kg and 4 kg capacities are available on the CPC-2000.
  • Page 6 CPC-2000 Technical Manual II Setting the Scale Functions YCO-1094-04-02...
  • Page 7: Calibrating The Scale

    CPC-2000 Technical Manual II Setting the Scale Functions Calibrating the Scale WARNING! Calibration should only be performed by a qualifi ed scale tech ni cian. Beware of the haz- ard of electric shock. Never work on powered circuits. Enter the technician setup mode, select “END” and press the...
  • Page 8: Troubleshooting

    CPC-2000 Technical Manual III Troubleshooting The scale can check various functions automatically. When a problem is detected, the fol low ing mes- sag es will be displayed. If you take the steps described below but the problem is not fi xed, the scale needs to be serviced by a qualifi...
  • Page 9: Wiring Diagram

    CPC-2000 Technical Manual IV Wiring Diagram DC JACK POWER BOARD LOAD CELL RECHARGEABLE BATTERY 6 V, 4Ah GREEN BLACK WHITE BLUE MAIN BOARD BAT DC JACK KEYPAD YCO-1094-04-02...
  • Page 10: Parts List

    CPC-2000 Technical Manual V Parts List YCO-1094-04-02...
  • Page 11 CPC-2000 Technical Manual VI Parts List ITEM PART NUMBER DESCRIPTION QTY. CHA-1595-000010 Stainless Steel Base CHA-1595-000020 Adjustable Feet CHA-1595-000030 Screw M4 x10 CHA-1595-000040 Rechargeable Battery CHA-1595-000050 Battery Holder CHA-1595-000060 Display / CPU Board CHA-1595-000070 Load Cell Assembly 4 lb / 2 kg...