Weinmann MEDUMAT Easy CPR Service And Repair Instructions page 58

Hide thumbs Also See for MEDUMAT Easy CPR:
Table of Contents

Advertisement

as per Servicing and Repair Instructions WM 67308 (Medumat Easy CPR)
 
3.9.
Function check of frequency setting
Frequency 14 ± 2 min
Frequency 10 ± 2 min
Frequency 30 ± 2 min
Function check of breath volume at 4.5 bar input pressure and 10 mbar back
3.10.
pressure
Breath volume 1140 ± 170 ml
Breath volume 1100 ± 170 ml
Breath volume 100 ± 20 ml
3.11.
Test O
concentration
2
O
concentration must be > 98 % by volume
2
3.12.
Function check of pressure limit
Pressure limit 20 ± 5 mbar
Pressure limit 45 ± 5 mbar
3.13.
Function check of venting valve without patient valve
The ventilator vents audibly
3.14.
Test the MEDUtrigger
LEDs on Medutrigger come on
LEDs on Medutrigger go out
3.15.
Test communication socket
The communication test runs without errors
(green 30 mbar LED comes on)
The communication test runs without errors
(red 60 mbar LED comes on)
3.16.
Check external condition
Housing not scratched and has no defects
Connecting thread undamaged and working smoothly
Knob inhibited to prevent unintentional adjustment
Elbow outlet is easy to rotate.
Geräte für Medizin GmbH+Co. KG, P.O. Box 54 02 68, D-22502 Hamburg, Germany, Fax +49 4054 702461,
Test record
-1
-1
-1
Telephone +49 4054 7020
Page 3 of 4
OK
NOK
-1
min
-1
min
-1
min
OK
NOK
ml
ml
ml
OK
NOK
OK
NOK
mbar
mbar
OK
NOK
OK
NOK
OK
NOK
OK
NOK
 