as per Servicing and Repair Instructions WM 67308 (Medumat Easy CPR)
3.9.
Function check of frequency setting
•
Frequency 14 ± 2 min
•
Frequency 10 ± 2 min
•
Frequency 30 ± 2 min
Function check of breath volume at 4.5 bar input pressure and 10 mbar back
3.10.
pressure
•
Breath volume 1140 ± 170 ml
•
Breath volume 1100 ± 170 ml
•
Breath volume 100 ± 20 ml
3.11.
Test O
concentration
2
•
O
concentration must be > 98 % by volume
2
3.12.
Function check of pressure limit
•
Pressure limit 20 ± 5 mbar
•
Pressure limit 45 ± 5 mbar
3.13.
Function check of venting valve without patient valve
•
The ventilator vents audibly
3.14.
Test the MEDUtrigger
•
LEDs on Medutrigger come on
•
LEDs on Medutrigger go out
3.15.
Test communication socket
•
The communication test runs without errors
(green 30 mbar LED comes on)
•
The communication test runs without errors
(red 60 mbar LED comes on)
3.16.
Check external condition
•
Housing not scratched and has no defects
•
Connecting thread undamaged and working smoothly
•
Knob inhibited to prevent unintentional adjustment
•
Elbow outlet is easy to rotate.
Geräte für Medizin GmbH+Co. KG, P.O. Box 54 02 68, D-22502 Hamburg, Germany, Fax +49 4054 702461,
Test record
-1
-1
-1
Telephone +49 4054 7020
Page 3 of 4
OK
NOK
-1
min
-1
min
-1
min
OK
NOK
ml
ml
ml
OK
NOK
OK
NOK
mbar
mbar
OK
NOK
OK
NOK
OK
NOK
OK
NOK