Close The Device - Weinmann MEDUMAT Easy CPR Service And Repair Instructions

Hide thumbs Also See for MEDUMAT Easy CPR:
Table of Contents

Advertisement

6.4 Close the device

Materials and tools required
ESD-protected workplace
Torque wrench 50 ± 5 Ncm,
Torque wrench 200 ± 10 Ncm,
Tubular hexagon box wrench, 10 mm
Special tool WM 22829 from special tool set WM 15349
Hex socket wrench, size 2.5
1. Place the two halves of the housing next to one
another with the outsides face down.
2. If the Y sensor tube was damaged: push tube
(shortest tube) and tube
59
Y sensor tube (tube set WM 15058) together
with Y connector
.
61
3. Put the battery contacts in the battery
compartment and screw them with the two
screws
and spring rings
48
Note:
The shortest tube 59 of the Y sensor tube may
only be pushed onto sensor S3003, otherwise it
may kink!
4. Push tube
(the shortest of the three tubes of
59
the Y sensor tube) carefully onto sensor S3003
onto sensor S3000 on the board.
and tube
60
5. Screw the long tube
Y sensor tube to the pneumatic block.
6. Screw the ground connection to the pneumatic
block tight using screw
washer
.
50
7. Connect long tube
58
Y sensor tube to Y connector
8. Screw the sensor tube to the pneumatic block.
9. Connect the connector for cable harness
terminal X1001 on the board.
10. Connect the connector for the MEDUtrigger to
terminal X1003 on the board.
32
Repair information and instructions
of the new
60
.
47
of the
58
and serrated lock
49
of the
.
61
to
53
60
59
61
S3003
S3000
48
59
60
61
X1003
X1001
58
47
58
Ground connection
53
50
49