7. Use a sharp object to push the blanking plug
of the interface connection out of lower part
66
of the housing
from the inside.
33
8. Take O-ring
out of lower part of the housing
40
.
33
9. Turn the special counter tool onto pressurized
gas connection
.
11
10. Clamp the special counter tool in a vise.
11. Turn the nut of the special counter tool against
the pressurized gas connection using an open-
ended wrench (size 22).
12. Unscrew elbow connector
ended wrench (size 13).
13. Use the special wrench (size 17) to slacken nut
and unscrew it completely.
39
14. Take metal piece
38
15. Put a new lower part of housing
pressurized gas connection
16. Push metal piece
38
housing onto the connection.
17. Screw nut
(size 17) on the inside of the
39
connection tight.
18. Fix elbow connector
19. Undo the nut of the special counter tool using
the open-ended wrench (size 22).
20. Open the vise.
21. Turn the special counter tool off pressurized
gas connection
.
11
22. Pass the cable for the MEDUtrigger socket
into the inside of the housing.
23. Tighten the nut on the MEDUtrigger socket
from the outside using the Allen key for
grooved nuts.
24. Pass the cable for communication socket
into the inside of the housing.
25. Tighten the nut on communication socket
26. Connect the cable for communication socket
to cable harness
54
27. Push blanking plug
66
connection from the outside.
28. Put O-ring
back in lower part of the housing
40
.
33
44
Repair information and instructions
using an open-
34
off upwards.
on
33
.
11
on the inside of the
on the connection.
34
55
55
54
.
54
.
53
into the interface
40
11
11
38
39
34
53
55
66
33
33
33
54

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wm 28140

Table of Contents