as per Servicing and Repair Instructions WM 67308 (Medumat Easy CPR)
Preparation
3.1.
Testing equipment (
3.2.
Preparation for the final check (
3.3.
Enter device data
Tests
3.4.
Check for leaks and check pressure display
3.4.1.
Check for leaks on inlet side
•
Drop in pressure is less than 0.2 bar/min
3.4.2.
Check for leaks in pressure measuring section
•
Drop in pressure is less than 2 bar/min
3.5.
Self-test after device is switched on
•
The self-test runs without errors
3.6.
Function check of controls (key test)
•
All the controls work perfectly
3.7.
Function check and alarms
3.7.1.
Test stenosis alarm
•
The stenosis alarm is correctly triggered
3.7.2.
Test alarm acknowledgement
•
The acoustic alarm is suppressed immediately, the visual alarm remains active
3.7.3.
Test disconnection alarm
•
The disconnection alarm is correctly triggered
3.7.4.
Test pressure alarm
•
The pressure alarm is correctly triggered
3.7.5.
Test demand mode LED
•
The LED must come on/flicker
3.7.6.
Test audio response
•
The audio response must be clear and intelligible.
3.8.
Battery voltage
3.8.1
3.0 V battery (board)
•
Battery voltage is between 3.0 and 3.2 V
3.8.1
3.6 V battery (battery compartment)
•
Battery voltage is between 3.4 and 3.7 V
Geräte für Medizin GmbH+Co. KG, P.O. Box 54 02 68, D-22502 Hamburg, Germany, Fax +49 4054 702461,
Test record
as per Servicing and Repair Instructions WM 67308)
as per Servicing and Repair Instructions WM 67308)
Telephone +49 4054 7020
Page 2 of 4
OK
NOK
OK
NOK
OK
NOK
OK
NOK
OK
NOK
V
V