Self-Test After The Device Is Switched On - Weinmann MEDUMAT Easy CPR Service And Repair Instructions

Hide thumbs Also See for MEDUMAT Easy CPR:
Table of Contents

Advertisement

3.4.2 Check for leaks in pressure measuring section
Ventilation
pressure indicator
1. Put T-tube connection with syringe (WM 15359) on the pressure measurement tube connection
2. Use the syringe to generate a pressure of 55 mbar ± 2 mbar on the test pressure gage.
Drop in pressure must be ≤ 2 mbar/min.
Requirement:

3.5 Self-test after the device is switched on

1. Provide a pressurized gas supply to MEDUMAT Easy CPR and to the flow meter.
2. Connect the patient circuit to MEDUMAT Easy CPR.
3. Open oxygen cylinder.
4. Switch on device.
The self-test is triggered, the ventilation mode LEDs light up once each in turn, the alarm
Requirement:
LEDs flash, the pressure indicator LEDs are run through 3x, a signal sounds and one sentence is issued
by the audio response.
14
Final check
AT
Test pressure gage
CP R
0, 9
10
11
15
Syringe
60
50
40
CP R
30
20
10
0
13
14
1,1
0,6
0,3
.
15