6.7 Changing the board

Tools and equipment required:
ESD-protected workplace,
Phillips screwdriver, size 1,
Tubular hexagon box spanner 10 mm,
Torque spanner 50 ±5 Ncm,
Torque spanner 200 ±10 Ncm,
Special tool WM 22829 from special tool set WM 15349,
Flat-nose pliers.
1. Open the device (see "6.3 Open the device" on page 30).
2. Release the ribbon cable from its clamp
: To do so, lift up the top part of the
X1000
clamp. Then you can pull out the cable.
3. Unscrew the two screws
board
from the upper part of the housing.
30
4. Unscrew the M3 nut from the sensor tube.
5. Place the new board
Make sure that the ribbon cable is not under the
board or jammed between it and the housing.
6. Attach the board
30
7. Place the ribbon cable in the clamp
the board and then press down the top part of
the clamp.
8. The screw of the sensor tube is secured with a
nut for transport. Remove the nut before screw-
ing the tube to the pneumatic block.
9. Close the device (see "6.4 Close the device"
on page 32).
10. Perform a final check (see "3. Final check" on
page 8).
36
Repair information and instructions