Guidance and Manufacturer's Declaration For Spot Vital
Signs LXi Radio
Emissions and Immunity Information
The Spot Vital Signs LXi is intended for use in the electromagnetic environment specified below. The customer or user of the Spot Vital Signs
LXi should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC 61000-3-3
The Spot Vital Signs LXi is intended for use in the electromagnetic environment specified below. The customer or user of the Spot Vital Signs
LXi should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Electrostatic discharge
± 6 kV contact
(ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast transient/
±2 kV for power supply
burst
lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential mode
IEC 61000-4-5
±2 kV common mode
Electromagnetic Emissions
Electromagnetic Environment - Guidance
The Spot Vital Signs LXi uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
The Spot Vital Signs LXi is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Electromagnetic Immunity
Compliance Level
Electromagnetic Environment - Guidance
± 6 kV contact
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
± 8 kV air
be at least 30%.
±2 kV for power supply
Mains power quality should be that of a typical commercial or
lines
hospital environment.
±1 kV for input/output
lines
±1 kV differential mode
Mains power quality should be that of a typical commercial or
hospital environment.
±2 kV common mode
Specifications
69