78
Troubleshooting
Table 3. Post Error Codes
Internal
Event Log
Code
Number
89
90
91
92
93
94
95
Table 4. SureTemp Plus Temperature Error Codes
Internal
Event Log
Code
Number
121
122
123
124
125
126
127
128
129
130
131
Displayed
Internal Event Log Description
Device Code
E39
No calibration signature.
E40
Bad NIBP calibration record.
E41
Stepper valve bad.
E42
8bit substitute for 16bit errors.
E43
Error log checksum error.
Err-E44
POST que full.
E44 E45
POST que full.
Displayed
Internal Event Log Description
Device Code
E0.4
Probe is over temperature (112° F /
44.4° C).
E0.1
Excessive heater energy.
23
Host interface error.
A^!
Device outside operating temperature
range.
Av!
Device outside operating temperature
range.
26
Invalid SureTemp algorithm.
b^
Battery voltage over the algorithm's
maximum value.
bv
Battery voltage under the algorithm's
minimum value.
29
Battery voltage not set.
30
Prediction algorithm not set.
31
Ambient temperature not set.
Welch Allyn Spot Vital Signs LXi
Corrective Action
General internal malfunction; contact
technical support.
General internal malfunction; contact
technical support.
General internal malfunction; contact
technical support.
Internal communications error.
Disconnect the battery and wait 5
minutes. Reconnect the battery, and then
see "Set the Date and Time
Configurations" in the Directions for Use.
General internal malfunction; contact
technical support.
General internal malfunction; contact
technical support.
General internal malfunction; contact
technical support.
Corrective Action
Correct probe tip conditions. If problem
persists, replace probe.
Try another predict cycle. If problem
persists, replace probe.
Try another predict cycle. If problem
persists, replace probe.
Decrease ambient temperature.
Increase ambient temperature.
Try another predict cycle. If problem
persists, replace probe.
Contact Technical Service.
Contact Technical Service.
Try another predict cycle. If problem
persists, replace probe.
Try another predict cycle. If problem
persists, replace probe.
Try another predict cycle. If problem
persists, replace probe.