104
Disassembly and Repair
8. Disconnect the red electrical connector and the gold ribbon connector. This is located
on the main board. Lift up the tab on the white cable retainer so cable can slide out
after it is unplugged.
9. Lift the flap on the beige connector (large ribbon cable) to remove the main board.
This releases the serial board cable from connector J108.
Note
FOR MODELS NUMBERS (45MT0, 45ME0) ONLY: With diagonal cutting pliers
cut jumper JP2 twice so that it is open. Cut it on each end on the dotted lines as
illustrated, so that it is open connection.
Welch Allyn Spot Vital Signs LXi
Gold Connector
Red Connector
J108