Table 4. SCC Module and Corresponding NI-DAQmx Measurement Type
Recommended NI-DAQmx
SCC Module
SCC-ACC01
Analog Input»Accelerometer
SCC-AI01
Analog Input»Voltage
SCC-AI02
Analog Input»Voltage
SCC-AI03
Analog Input»Voltage
SCC-AI04
Analog Input»Voltage
SCC-AI05
Analog Input»Voltage
SCC-AI06
Analog Input»Voltage
SCC-AI07
Analog Input»Voltage
SCC-AI13
Analog Input»Voltage
SCC-AI14
Analog Input»Voltage
SCC-A10
Analog Input»Voltage
SCC-AO10
Analog Output»Voltage
SCC-CI20
Analog Input»Current
SCC-CO20
Analog Output»Current
SCC-DI01
Digital Input
SCC-DO01
Digital Output
SCC-FT01
Any measurement type
SCC-FV01
Analog Input»Frequency
SCC-LP01
Analog Input»Voltage
SCC-LP02
Analog Input»Voltage
SCC-LP03
Analog Input»Voltage
SCC-LP04
Analog Input»Voltage
SCC-RLY01
Digital Output
SCC-RTD01
Analog Input»Temperature»RTD
SCC-SG01
Analog Input»Strain Gage
SCC Quick Start Guide
Measurement Type
Recommended
Parameter Settings
Current Excitation Value = 4 mA,
Excitation Source = Internal
Sensitivity = #
Max/Min = ±42 V
Max/Min = ±20 V
Max/Min = ±10 V
Max/Min = ±5 V
Max/Min = ±1 V
Max/Min = ±100 mV
Max/Min = ±50 mV
Max/Min = ±10 V
Max/Min = ±5 V
Max/Min = ±100 V
Max/Min = ±10 V
Max = 20 mA, Min = 0 mA
Max = 20 mA, Min = 0 mA
Use default
Use default
Based on your application
Use default
Max/Min = ±10 V
Max/Min = ±10 V
Max/Min = ±10 V
Max/Min = ±10 V
Use default
Current Excitation Value = 1 mA
Excitation Source = Internal
RTD Type = #
R
= #
0
Resistance Configuration = #
Strain Configuration = Quarter Bridge I
Gage Resistance = 120 Ω
Gage Factor = #