Fluke RUSKA 2465A-754 User Manual page 33

Gas piston gauge
Table of Contents

Advertisement

Figure 5-5. Handling the Low Range Piston and Cylinder
The Low Range Cylinder (2460-5-1) and Low Range Piston Assembly (2460-55) should
then be placed in the gauge after being cleaned according to the instructions in Chapter 6.
See Figure 5-6.
The O-Ring groove on the cylinder should be in the downward
position when the cylinder is placed in the gauge.
Figure 5-6. Low Range Piston and Cylinder Showing O-Ring Groove
When screwing the Retaining Cap (2460-4-7) onto the cylinder, a slight upward force
may be necessary on the piston to properly seat the Piston Retainer (2460-5-5) into the
inner recess of the Retaining Cap. See Figure 5-7.
W Caution
Operation
Low Range Piston Assembly
glg10.eps
glg11.eps
5