Temperature Range; Humidity Range; Pressure Medium; Pressure - Fluke RUSKA 2465A-754 User Manual

Gas piston gauge
Table of Contents

Advertisement

Temperature Range

Operating
Storage

Humidity Range

Operating
Storage

Pressure Medium

Clean dry gas, Nitrogen or equivalent, regulated to a Pressure Compatible with each
particular piston/cylinder assembly. Dew Point of less than or equal to -51 ºC (-60 ºF).

Pressure

Maximum Working Pressure
2465
With High Range Piston/Cylinder
With Upper Mid Range Piston
With Low Range Piston/Cylinder
With Lower Range Piston/Cylinder
2468
With Upper Mid Range Piston
With Lower Mid Range Piston/Cylinder
With Low Range Piston/Cylinder

Power Requirements

2465A-754 (DWG base): 15 watts, 115 VAC or 230 VAC; 50 or 60 Hz
(1)
Negative gauge pressure limited by local barometric pressure.
(2)
Absolute mode uncertainty higher due to reference pressure sensor.
(3)
Approximate 95% level of confidence (Refer to Calibration Report.)
(4)
Low mass carrier and plate may be of other non-magnetic material.
(5)
Limited by dissimilar metals of Low range Piston/Cylinder.
18 ºC to 28 ºC
(5)
-40 ºC to 70 ºC
20% to 75% non-condensing
0% to 90% non-condensing
7000 kPa
3500 kPa
700 kPa
172 kPa
7000 kPa
1380 kPa
345 kPa
Specifications
Temperature Range
(1000 PSIG)
(500 PSIG)
(100 PSIG)
(25 PSIG)
(1000 PSIG)
(200 PSIG)
(50 PSIG)
1
1-3