Thermo Scientific FH100D Operating Manual page 29

Dispensing pump systems
Table of Contents

Advertisement

Volume Dispense
Mode Operation
(continued)
Thermo Scientific
7. Setting Batch Size: Use the directional pad and ENTER key to select
the Batch Count icon from the operation screen (see position F,
Figure 3-11). From Batch Count screen use the UP/DOWN arrows to
select batch size. Press ENTER to validate batch size. When set to zero
(0) the drive will run for an infinite number of cycles and the
symbol is displayed. Select EXIT to return to drive operation screen.
Batch count may be reset to zero from BATCH COUNT screen
by selecting RESET.
8. Operating the Drive: Press the START/STOP key to operate the drive
continuously at the settings selected and displayed on the screen. Press
again to pause or stop the drive. Drive will automatically stop once
batch is complete.
9. Reset Batch Count: To reset a BATCH COUNT to zero, use the
directional pad and the ENTER key to select the BATCH COUNT
icon (see position F, Figure 3-11). In the BATCH COUNT screen,
select RESET and press the ENTER key to reset the batch count.
Select EXIT to return to the COPY MODE OPERATION screen.
Table 3-2. Typical maximum single dispense volume in VOLUME DISPENSE
MODE.
TUBING SIZE
13 (0.8 mm)
14 (1.6 mm)
16 (3.2 mm)
15/25 (4.8 mm)
17/24 (6.4 mm)
18/35 (8.0 mm)
36 (9.5 mm)
NOTE: The drive is limited to a maximum number of revolutions that can
be stored accurately in VOLUME DISPENSE MODE. The maximum
number of revolutions safely stored is 5,461. See the chart for the
equivalent maximum volume by tube size. If a volume dispense value
exceeding this is entered inaccurate dispenses will result. If greater volumes
are required change to TIME DISPENSE MODE or increase the tubing
size used.
VOL/REV mL
0.1
0.38
1.38
3.0
5.0
7.3
10.0
FH100D/FH100DX Dispensing Pump Operating Manual
Section 3
Operation
Max Disp Vol mL
500
2075
7500
16000
27000
39500
54000
3-19