Section 3
Operation
Copy Dispense Mode
Operation (continued)
3-14
FH100D/FH100DX Dispensing Pump Operating Manual
8. Operating Drive: Press the START/STOP key to operate the drive at
the settings selected and displayed on the screen. Press again to pause
or stop the drive. Drive will automatically stop once batch is complete.
9. Reset Batch Count: To reset a BATCH COUNT to zero, use the
directional pad and the ENTER key to select the BATCH COUNT
icon (see position F, Figure 3-7). In the BATCH COUNT screen,
select RESET and press the ENTER key to reset the batch count.
Select EXIT to return to the Copy Mode Operation screen.
Table 3-1. Typical maximum single dispense volume in COPY MODE.
TUBING SIZE
13 (0.8 mm)
14 (1.6 mm)
16 (3.2 mm)
15/25 (4.8 mm)
17/24 (6.4 mm)
18/35 (8.0 mm)
36 (9.5 mm)
NOTE: The drive is limited to a maximum number of revolutions that can
be stored accurately in COPY DISPENSE MODE. The maximum num-
ber of revolutions safely stored is 5,461. At full speed the maximum time
for a copy is approximately 13-1/2 minutes. If a copy dispense exceeding
this is created inaccurate dispenses will result. If greater copy volumes are
required please change to TIME DISPENSE MODE or increase the tub-
ing size used.
VOL/REV mL
0.1
0.38
1.38
3.0
5.0
7.3
10.0
Max Disp Vol mL
500
2075
7500
16000
27000
39500
54000